| Pulmonetic LTV800 Ventilator | |||
Basic Description | |||
| Manufacturer: | Pulmonetic Systems | ||
| Ventilator Type: | Positive Pressure Negative Pressure Invasive | ||
| Ventilator Class: | Transport Subacute Home | ||
| Patient Population Range: | Pediatric Adult Acute Long-term (chronic) | ||
Description | |||
| Brief Description: | The LTV800 is designed for the patient who requires volume ventilation, particularly for homecare application. | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- ASSIST/CONTROL; SIMV Pressure-controlled modes --- CPAP | ||
| Modes--Non-Invasive: | Volume-controlled modes: ASSIST/CONTROL; SIMV; SPONT. | ||
| Operating Range of Controls: | Breath Rate: 0 to 80 BPM; Tidal Volume: 50 to 2000 ml; Inspiratory Time: 0.3 to 9.9 sec.; Sensitivity: Off, 2 to 20 cmH2O; PEEP/CPAP: 0 to 20 cmH2O | ||
| Graphic Display: | LTM (Laptop Monitor) - Real time waveform of Pressure, Flow and Tidal Volume with Freeze and cursors; Real time loops of Pressure/Volume and Flow/Volume with Reference Loop and up to 4 historical loops; Provides trending for 24 hours on f, Spf, Vte, VE, PIP, MAP, PEEP, Static Compliance and Auto PEEP; Displays ventilator settings and monitored values; Communicates with Patient Data Systems; Available print on all screens; Touch Screen color panel that can be custom configured | ||
| Size: | 3" x 10" x 12" (8 x 23 x 30 cm) | ||
| Weight: | 12.4 lbs (5.66 kg) | ||
Characteristics (Non - Technical) | |||
| Approvals: | ETL/cETL Listed, IEC 601.1, ISO 9001/EN 46001 | ||
Specifications | |||
| Front Panel Controls: | Power: On/Standby; Ventilation Modes: Control, Assist/Control, SIMV, CPAP; Breath Types: Volume Control, Spontaneous; Breath Rate; Tidal Volume; Inspiratory Time; Sensitivity; PEEP/CPAP; Control Lock: Easy or Hard unlock methods; Manual Breath: 1 x current settings; | ||
| Alarm Controls: | Apnea Interval: 10-60 sec; High Pressure Limit: 5 - 100 cmH2O; Low Peak Pressure: Off, 1-60 cmH2O; Alarm Volume; Alarm Silence/Reset: 60 sec; Ventilator Inoperative | ||
| Non-Adjustable Alarms: | Disconnect; Low & Lost External Power; Low & Empty Internal Battery; | ||
| AC Power and Battery Indicators: | Red, Green, Yellow LEDs | ||