| Newport HT50 Ventilator | |||
Basic Description | |||
| Manufacturer: | Newport Medical Instruments | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) Transport Subacute Home CPAP | ||
| Patient Population Range: | Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- A/CMV
SIMV
Spontaneous (SPONT) Pressure-controlled modes --- A/CMV SIMV Spontaneous (SPONT) | ||
| Operating Range of Controls: | Volume Control (Tidal Volume): 100 to 2,200 mL; Frequency: 1 to 99 b/min; Inspiratory Time: 0.1 to 3.0 sec; PEEP/CPAP: 0 to 30 cmH2O / 0 to 30 mbar; Pressure Support: 0 to 60 cmH2O / 0 to 30 mbar; Flow: 6 to 100 L/min; I:E Ratio: 1:99 to 3:1; Pressure Control (Target Pressure): 5 to 60 cmH2O / 5 to 60 mbar; Ptrig (Sensitivity): -9.9 to 0 cmH2O / -9.9 to 0 mbar; Panel Lock: ON/OFF; Manual Inflation: 3 sec maximum (single operator controlled breath); FIO2 (optional): .21 to 1.00; Humidifier (optional): 19 to 39 degrees C | ||
| Size: | 10.63 x 7.87 x 10.24 in (27 x 20 x 26 cm) width/depth/height | ||
| Weight: | 15 lbs (7.5 kg) w/o humidifier; 17.2 lbs (7.8 kg) with humidifier | ||
Characteristics (Non - Technical) | |||
| Approvals: | CE, AMTAC, MEDIQA | ||
Specifications | |||
| Power Requirements: | 100-240 VAC, max. 2A 50/60/400 Hz 12-30 VDC, max. 12A | ||
Related Info | |||