| Sechrist Millennium Infant Pediatric Ventilator | |||
Basic Description | |||
| Manufacturer: | Sechrist Industries | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) Subacute | ||
| Patient Population Range: | Neonatal Pediatric | ||
Characteristics | |||
| Modes--Invasive: | General modes --- AC, SIMV/IMV, CPAP, CPAP with backup Ventilation, Standby | ||
| Operating Range of Controls: | Flow: 2-32 LPM; flush 40 LPM; Inspiratory Time: 0.1 to 3.0 seconds; I: E Ratio: 4:1 to 1:10 (settable); Rate: 2-150 breaths per minute; Expiratory Pressure: 0 to 20 cmH20; Inspiratory Pressure: 5 to 70 cmH20; Waveform: Tapered to square; FiO2: 0.21 to 1.00; Manual breath: Electronically controlled, based on the Inspiratory pressure, Inspiratory time and waveform control settings | ||
| Size: | 13.5"H x 11" x 9.5 " | ||
| Weight: | 21 lbs. | ||
| Other Characteristics: | Also monitors the calculated I:E ratio | ||
Specifications | |||
| Power Requirements: | 110-120 VAC 50/60 Hz (220-240 VAC 50/60 Hz) | ||
| Alarms: | High Inspiratory Pressure; Low/Prolonged Inspiratory Pressure; Low/PEEP/Baseline Pressure; Apnea; High Respiratory Rate; Loss/Low Electrical Power; System Error/Vent. Inop; Loss of Inlet Gas Source; Alarm Delay: 3 to 15 Seconds; Alarm Mute: 30 Seconds; Safety Pressure Relief Valve: 15 to 85 cmH20 | ||
| Other Specifications: | Gas Consumption: 14 LPM above the continuous flow setting | ||
Monitored Variables : | |||
| Pressure | Inspiratory Pressure, PEEP/Baseline Pressure, Mean Airway Pressure, Proximal Airway Pressure Bar Graph | ||
| Respiratory Rate | Measured | ||
Related Info | |||