Sechrist Millennium Infant Pediatric Ventilator
   

Basic Description

 
 
Manufacturer:  Sechrist Industries
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Subacute
Patient Population Range:  Neonatal
Pediatric
 

Characteristics

 
Modes--Invasive:  General modes --- AC, SIMV/IMV, CPAP, CPAP with backup Ventilation, Standby
Operating Range of Controls:  Flow: 2-32 LPM; flush 40 LPM; Inspiratory Time: 0.1 to 3.0 seconds; I: E Ratio: 4:1 to 1:10 (settable); Rate: 2-150 breaths per minute; Expiratory Pressure: 0 to 20 cmH20; Inspiratory Pressure: 5 to 70 cmH20; Waveform: Tapered to square; FiO2: 0.21 to 1.00; Manual breath: Electronically controlled, based on the Inspiratory pressure, Inspiratory time and waveform control settings
Size:  13.5"H x 11" x 9.5 "
Weight:  21 lbs.
Other Characteristics:  Also monitors the calculated I:E ratio

Specifications

Power Requirements:  110-120 VAC 50/60 Hz (220-240 VAC 50/60 Hz)
Alarms:  High Inspiratory Pressure; Low/Prolonged Inspiratory Pressure; Low/PEEP/Baseline Pressure; Apnea; High Respiratory Rate; Loss/Low Electrical Power; System Error/Vent. Inop; Loss of Inlet Gas Source; Alarm Delay: 3 to 15 Seconds; Alarm Mute: 30 Seconds; Safety Pressure Relief Valve: 15 to 85 cmH20
Other Specifications:  Gas Consumption: 14 LPM above the continuous flow setting

Monitored Variables :

Pressure Inspiratory Pressure, PEEP/Baseline Pressure, Mean Airway Pressure, Proximal Airway Pressure Bar Graph
Respiratory Rate Measured

Related Info 