| Model: | Puritan-Bennett Option 80, Pressure Control Ventilation (PCV) | ||
| Manufacturer: | Puritan Bennett Corporation | ||
| Accessory Type: | Supplies (Miscellaneous) | ||
| For use with: | Puritan-Bennett 7200ae Ventilator Puritan-Bennett 7200spe Ventilator | ||
Characteristics | |||
| Operating Range of Controls: | Pressure Control Ventilation Parameters: Inspiratory pressure (above PEEP): 5 to 100 cmH2O; Inspiratory Time: 0.2 to 9.9 seconds; I:E ratio: Inspiratory component between 4 and 1, expiratory component between 9 and 1; Respiratory rate: 0.5 to 70 bpm; Oxygen percentage: 21% to 100%; High pressure limit: 10 to 120 cmH2O; Pressure Control Ventilation Apnea Parameters: Apnea interval: 10 to 60 seconds; Inspiratory pressure (above PEEP): 5 to 100 cmH2O; Inspiratory Time: 0.2 to 9.9 seconds; I:E ratio: Inspiratory component between 4 and 1, expiratory component between 9 and 1; Respiratory rate: 0.5 to 70 bpm; Oxygen percentage: 21% to 100%; High pressure limit: 10 to 120 cmH2O | ||
| Other Characteristics: | Supports I:E up to 4:1; Delivers breaths based on set inspiratory time or I:E ratio; Holds I:E ratio or inspiratory time constant with respiratory rate changes; For mandatory breaths in CMV, SIMV, and CPAP modes; Compatible with Pressure Support, Respiratory Mechanics, and Flow-by; Provides separate parameters for pressure-controlled apnea ventilation | ||