Bird 8400ST Ventilator
(Discontinued)
   

Basic Description

 
 
Manufacturer:  Bird Products Corporation
Ventilator Type:  Positive Pressure
Invasive
Conventional
 

Characteristics

 
Modes--Invasive:  Volume-controlled modes --- CMV, IMV/SIMV
Pressure-controlled modes --- CMV, IMV/SIMV (spontaneous), CPAP
Flow-controlled modes --- ASSIST/CONTROL, SIMV (mandatory)
Supplemental modes --- Sigh
Waveforms:  Pressure: Rectangular; Flow: Rectangular, Descending Ramp
Operating Range of Controls:  Tidal Volume: 50 to 2000 ml; Inspiratory Pressure: 5 to 100 cmH20; Peak Flow: 10 to 120 Ipm; Tidal Volume: 50 to 2000 ml; Maximum Inspiratory Pressure Threshold: 0 to 140 cmH2O; Peak Flow: 10 to 120 Ipm; Breath Rate: 0 to 80 bpm; PEEP/CPAP or baseline pressure: 0 to 30 cmH2O; Sensitivity: Pressure Triggered: Flow Triggered: 1 to 10 Ipm, Off; Baseline Flow: 10 Ipm; Pressure Support: 1 to 50 cmH20 above PEEP/CPAP, Off; Sigh: On / Off, I Sigh / 100 Breaths; Sigh Volume: 1.5 x Set Tidal Volume; Oxygen%: (with blender) 21 to 100%
Triggering:  Time, pressure, flow
Other Characteristics:  I:E ratio is also displayed.

Specifications

Front Panel Controls:  Mode: Assist; SIMV; Control; CPAP; Tidal Volume; Peak Flow; Breath Rate; PEEP CPAP; Sensitivity; Pressure Support
Power Requirements:  Electrical: 120 VAC, 60 Hz or optional European voltages; Pneumatic: 30 to 70 psig
Alarms:  Alarm Silence: 60 Sec.; Alarm Reset: Active in all modes; Alarm Volume: Adjustable 74 to 84 dB; Remote Alarm Output Jack: 1/4" 24 Volts 112 amps Maximum Load
Alarm Controls:  High Inspiratory Pressure Limit: 1 to 140 cmH20; Sigh Pressure Limit: 1.5 x Set; Low Peak Pressure: 2 to 140 cmH20, Off; Low PEEP / CPAP: -20 to +30 cmH20; Low Minute Volume: 0 to 99.9 L; High Breath Rate: 3 to 150 bpm; Apnea Interval: 10 to 60 sec.; Back up Breath Rate: 0 to 80 bpm; I:E Ratio: Inspiratory time > Expiratory Time
Non-Adjustable Alarms:  Low Inlet Gas: 17 PSIG (1.19 kg cm2); Ventilator Inoperative: Red Indicator, Flashing Display; Circuit alarm: circuit occluded or kinked, blocked airway sensing port, or pressure transducer failure

Monitored Variables :

Pressure gauge
Respiratory Rate Total (spontaneous + mandatory)
Tidal Volume
Minute Volume