| Model: | Siemens CO2 Analyzer Module for Servo Screen 390 |
| |
| Manufacturer: | MAQUET GmbH & Co. KG |
| Accessory Type: | Monitors |
| For use with: | Siemens Servo 300 Ventilator |
| |
Characteristics |
| Graphic Display: | The CO2 analyzer module presents the following on the Servo Screen 390:
CO2 concentration - Time waveform (capnogram) depicting the variation in airway CO2 level during the patient's respiratory cycle;
CO2 concentration - Expired volume waveform (single breath carbon dioxide waveform);
End-tidal CO2 concentration (etCO2) |
| Size: | CO2 analyzer module: 150 x 93 x 70 mm;
Sensor: 32 x 42 x 22 mm |
| Weight: | CO2 analyzer module: 0.45 kg;
Sensor: <18 g;
Adapter: 10 g |
| Other Characteristics: | Also monitors Airway Dead Space (VDaw), End Tidal CO2 concentration (etCO2%), Capnogram (CO2conc-Time waveform, SBCO2 (CO2conc-Exp. volume waveform, single breath carbon dioxide waveform) and Alveolar Ventilation (VA).
Oxygen concentration and barometric pressure compensation are automatic from values supplied by the SV300 system. |
Characteristics (Non - Technical) |
| Other Comments: | CO2 analyzer module kit for attachment to the Siemens Servo Screen 390 with software memory card, version 2.0 or higher:
etCO2 Capnostat III sensor with 2.4 m cable;
etCO2 airway adapter, adult (reusable);
CO2 analyser module;
Holder for CO2 analyser module;
Calibration and reference cell;
Cable clips, 5 each;
24 V power branch cable, 1.2 m;
Operating manual (six language versions);
Neonatal airway adapter (reusable) can be ordered separately |
|
Specifications |
| Power Requirements: | CO2 analyzer module:
Supply voltage: Powered from the SV300;
Power:
Warmup: 7 W;
During normal working conditions: 4 W;
Sensor: Powered from the CO2 analyzer module |
| Languages: | For the software upgrade kit for the SS390, version 2.0, (software memory card and operating manual): English, German, French, Spanish, Italian, Swedish |
| Non-Adjustable Alarms: | etCO2:
%: minimum 0.5, maximum 20;
mm Hg: minimum 4, maximum 100;
kPa: minimum 0.5, maximum 14 |
| Other Specifications: | Operating and Storage Conditions:
CO2 analyzer module, including sensor:
Operating: 10° to 40°C, 20% to 90% relative humidity, non-condensing;
Storage: -20° to 50°C, 20% to 95% relative humidity, with packaging;
CO2 analyzer module, including adaptor:
Operating: 700 to 1060 hPa (525 to 795 mmHg) atmospheric pressure;
Storage: 500 to 1060 hPa (325 to 795 mmHg) atmospheric pressure
Warm-up time:
30 seconds to initial CO2 indication;
maximum 5 minutes to full specification
Airway adapter deadspace:
Adult: <5 cm3;
Neonate: <0.5 cm3 |
Monitored Variables : |
| Tidal Volume | CO2 Tidal Production; Alveolar Tidal Volume (VTA) |
| Minute Volume | CO2 Minute Production |
Related Info |
Distributor | Educator |