| Bird V.I.P. Bird Sterling | |||
Basic Description | |||
| Manufacturer: | Bird Products Corporation | ||
| Patient Population Range: | Neonatal Pediatric | ||
Characteristics | |||
| Operating Range of Controls: | Apnea Switch 10 to 60 sec.; Assist Sensitivity Pressure Triggered 1 to 20 cmH2O, OFF; Assist Sensitivity Flow Triggered w/Inf. Sensor (all modes) 0.2 to 3 lpm, OFF; Assist Sensitivity Flow Triggered w/Ped. Sensor (no TCPL) 1.0 to 5.0 lpm, OFF; Bias Flow w/Infant Flow Sensor 3 lpm (except TCPL); Bias Flow w/Pediatric Flow Sensor OFF, 5 lpm; Flow 3 to 120 lpm (Volume and VAPS), 3 to 40 lpm (TCPL, Insp. Flow), 3 to 15 lpm (TCPL, Exp. Bias Flow); Inspiratory Pressure 3 to 80 cmH2O; Inspiratory Time 0.10 to 3.00 sec.; O2 concentration 21% to 100%, PEEP/CPAP 0 to 24 cmH2O; Pressure Relief Valve 0 to 130 cmH2O; Pressure Support OFF, 1 to 50 cmH2O; Rate 1 to 150 bpm; Termination Sensitivity 5 to 25% of peak flow, OFF; Tidal Volume 10 ml to 1200 ml | ||
| Graphic Display: | Waveform display of Pressure, Flow, and Volume with Freeze and cursor functions, Pressure-Volume and Flow-Volume Loops with capabilities for storing Reference Loops, Mechanics calculations for Dynamic and Static Compliance and Resistance, Trending of all parameters for up to 24 hours | ||
Specifications | |||
| Alarm Controls: | High Breath Rate Alarm OFF/ 0 to 300 bpm; High Pressure Alarm 3 to 120 cmH2O; High Tidal Volume OFF/2 ml to 2 L; Low Minute Volume Alarm: OFF, 0.01 to 30 L; Low Peak Pressure: OFF/ 3 to 120 cm H2O; Pressure Support/ VAPS Time Limit: O.10 to 3.00 sec. | ||
| Non-Adjustable Alarms: | Blender Input Gas Alarm:(Air & Oxygen Difference > +/- 20 PSIG); High/Prolonged Pressure Alarm: PEEP/CPAP + 6 cmH2O; Low Inlet Gas Pressure Alarm: < 22.5 PSIG or > 27.4 PSIG; Low Peep/CPAP Pressure Alarm: Fixed at 3 cmH2O below set PEEP/CPAP | ||
| Other Specifications: | V.I.P. Bird Sterling comes with Infant and Pediatric Flow Sensors, Stand, and Oxygen Analyser, and is available as English 120v, English 220v, English 240v, English (UK), English (Canada), English (Scandinavia), German, Italian, Spanish, Spanish (220v), French, and Japanese. The Bird Graphics Monitor is available separately and interfaces with any Bird ventilator. | ||
Monitored Variables : | |||
| Pressure | Airway Pressure Manometer -20 to 140 cmH2O (minimum); Mean Airway Pressure (MAP) 1 to 120 cm H2O; Peak Inspiratory Pressure 1 to 130 cmH2O; Positive End Expiratory Pressure (PEEP) 0 to 24 cmH2O | ||
| Respiratory Rate | 0 to 250 bpm | ||
| Tidal Volume | 0 to 9999 ml w/flow sensor; OFF w/no flow sensor | ||
| Minute Volume | 0.0 to 99.9 | ||
Related Info | |||