Pulmonetic LTV950 Ventilator
   

Basic Description

 
 
Manufacturer:  Pulmonetic Systems
Ventilator Type:  Positive Pressure
Invasive
Noninvasive
Conventional
Ventilator Class:  Transport
Subacute
Home
CPAP
Patient Population Range:  Pediatric
Adult
Acute
Long-term (chronic)
 

Characteristics

 
Modes--Invasive:  Volume-controlled modes --- ASSIST/CONTROL; SIMV (PSV);
Pressure-controlled modes --- ASSIST/CONTROL; SIMV (PSV); CPAP
Modes--Non-Invasive:  Volume-controlled modes: ASSIST/CONTROL; SIMV; SPONT.; Pressure-controlled modes: ASSIST/CONTROL; SIMV; SPONT.
Operating Range of Controls:  Breath Rate: 0 - 80 breaths per minute; Tidal Volume: 50 - 2000 ml; Inspiratory Time: 0.3 - 9.9 seconds (100 lpm); PC/PS/Spontaneous Flow: >/= 140 lpm; Pressure Support: Off, 1 - 60 cmH2O; Sensitivity: Off, 1 - 9 lpm; Bias Flow: 10 lpm during exhalation; PEEP/CPAP: 0 - 20 cmH2O
Graphic Display:  Real time waveform of Pressure, Flow and Tidal Volume with Freeze and cursors; Real time loops of Pressure/Volume and Flow/Volume with Reference Loop and up to 4 historical loops; Provides trending for 24 hours on f, Spf, Vte, VE, PIP, MAP, PEEP, Static Compliance and Auto PEEP; Displays ventilator settings and monitored values; Communicates with Patient Data Systems; Available print on all screens; Touch Screen color panel that can be custom configured
Size:  3"H x 10"W x 12"D (8cm x 23cm x 30cm)
Weight:  12.4 lb (5.66 kg)
Other Characteristics:  PC/PS Variable Rise Time: 0.1 to 0.9 sec.; PC / PS Variable Flow Termination: 10% to 40%; PS Variable Time Termination: 1.0 to 3.0 sec.; Pressure Control Flow Termination: On/Off; Remote Alarm Connection; Has a cigarette lighter adapter

Characteristics (Non - Technical)

Approvals:  ETL, cETL Listed;IEC 601.1, ISO 9001/EN46001 Registration;BSI Registered, FM 45841
Other Comments:  20,000 hour service plan: 2-10,000 hour maintenance; 30,000 hour service plan: 2-10,000 hour and 1-30,000 hour maintenance

Specifications

Front Panel Controls:  Power: On/Standby; Ventilation Modes: Control, Assist/Control, SIMV, CPAP; Breath Types: Volume Control, Pressure Control, Pressure Support, Spontaneous; Breath Rate; Tidal Volume; Inspiratory Time; PC/PS/Spontaneous Flow; Pressure Support; Sensitivity; Bias Flow; PEEP/CPAP; Control Lock: Easy or Hard unlock methods; Manual Breath: 1 x current settings
Power Requirements:  Electrical Requirements: 11 to 15 VDC; 90 to 250 VAC; 47 to 63 Hz; Pneumatic Requirements; Low Pressure Source: < 80 lpm Internal Battery: 60 mins. External Battery: 3 Hr or 9 Hr. Source Flow Oxygen (Medical Grade): 100 L/min
Languages:  English, French, German, Spanish, Italian, Swedish, Danish, Japanese
Alarms:  High Pressure Audible Alarm Delay: No Delay, 1 Breath, 2 Breaths Alarm Silence/Reset: 60 seconds; Ventilator Inoperative: Red LED; Disc/Sense; Low and Lost External Power; Low and Empty Internal Battery; High and Low Oxygen Inlet Pressure
Alarm Controls:  Apnea Interval: 10 to 60 seconds; High Pressure Limit: 5 to 100 cmH2O; Low Peak Pressure: Off, 1 to 60 cmH2O; Low Minute Volume: Off, 0.1 to 99 liters; Alarm Volume: 60 to 85 dBA at 1 meter
Non-Adjustable Alarms:  Disconnect; Low & Lost External Power; Low & Empty Internal Battery; High & Low Oxygen Inlet Pressure; Hardware Fault; Transducer Fault; Vent Reset; Setting Defaults
AC Power and Battery Indicators:  Red, Green, Yellow LEDs
Other Specifications:  Oxygen Source: Low pressure bleed in; Bias Flow: 10 lpm during exhalation; Over Pressure Relief: 110 cmH2O; Sub-Ambient Relief: Pressure Drop: -5 cmH2O; Control Lock: Easy or Hard Unlock methods; Manual Breath: single volume or pressure control breath NPPV Mode: Alarm package limited to: High Pressure, Disconnect, Defaults, Power Lost, Batter Low, Battery Empty, Apnea, Ventilator Inoperative Extended Features: NPPV Mode; Apnea Back-up Ventilation; Pressure Control and Pressure Support Variable Rise Time; Pressure Control and Pressure Support Variable Flow Termination Criteria; Pressure Support Variable Time Termination; Leak Compensation Other Monitored Variables: IE Ratio: 99:1 - 1:99; Calculated Peak Flow: 10 - 100 lpm; Patient Effort: Green LED; Monitor Scrolling

Monitored Variables :

Pressure Peak Inspiratory Pressure: 0 - 120 cmH2O, Mean Airway Pressure: 0 - 99 cmH2O, PEEP: 0 - 99 cmH2O, Airway Pressure Display: -10 to 108 cmH2O
Respiratory Rate Total Breath Rate: 0 - 250 bpm
Tidal Volume Exhalation Tidal Volume: 0 - 4000 ml
Minute Volume Total Minute Volume: 0 - 99.9 liters

Related Info 

Distributor |Educator | Manufacturer |Rental Agent | Service Center |
Parts Available | Product Markets | Service Centres |

Distributor:

Pulmonetic Systems
930 S Mt. Vernon Ave Suite 100
Colton, CA



Manufacturer:

Pulmonetic Systems
930 S Mt. Vernon Ave Suite 100
Colton, CA



Service Center:

Mobile Medical Maintenance Co.
10555 Woodbrook Dr.
Cement City, MI