| Pulmonetic LTV950 Ventilator | |||
Basic Description | |||
| Manufacturer: | Pulmonetic Systems | ||
| Ventilator Type: | Positive Pressure Invasive Noninvasive Conventional | ||
| Ventilator Class: | Transport Subacute Home CPAP | ||
| Patient Population Range: | Pediatric Adult Acute Long-term (chronic) | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- ASSIST/CONTROL; SIMV (PSV); Pressure-controlled modes --- ASSIST/CONTROL; SIMV (PSV); CPAP | ||
| Modes--Non-Invasive: | Volume-controlled modes: ASSIST/CONTROL; SIMV; SPONT.; Pressure-controlled modes: ASSIST/CONTROL; SIMV; SPONT. | ||
| Operating Range of Controls: | Breath Rate: 0 - 80 breaths per minute; Tidal Volume: 50 - 2000 ml; Inspiratory Time: 0.3 - 9.9 seconds (100 lpm); PC/PS/Spontaneous Flow: >/= 140 lpm; Pressure Support: Off, 1 - 60 cmH2O; Sensitivity: Off, 1 - 9 lpm; Bias Flow: 10 lpm during exhalation; PEEP/CPAP: 0 - 20 cmH2O | ||
| Graphic Display: | Real time waveform of Pressure, Flow and Tidal Volume with Freeze and cursors; Real time loops of Pressure/Volume and Flow/Volume with Reference Loop and up to 4 historical loops; Provides trending for 24 hours on f, Spf, Vte, VE, PIP, MAP, PEEP, Static Compliance and Auto PEEP; Displays ventilator settings and monitored values; Communicates with Patient Data Systems; Available print on all screens; Touch Screen color panel that can be custom configured | ||
| Size: | 3"H x 10"W x 12"D (8cm x 23cm x 30cm) | ||
| Weight: | 12.4 lb (5.66 kg) | ||
| Other Characteristics: | PC/PS Variable Rise Time: 0.1 to 0.9 sec.; PC / PS Variable Flow Termination: 10% to 40%; PS Variable Time Termination: 1.0 to 3.0 sec.; Pressure Control Flow Termination: On/Off; Remote Alarm Connection; Has a cigarette lighter adapter | ||
Characteristics (Non - Technical) | |||
| Approvals: | ETL, cETL Listed;IEC 601.1, ISO 9001/EN46001 Registration;BSI Registered, FM 45841 | ||
| Other Comments: | 20,000 hour service plan: 2-10,000 hour maintenance; 30,000 hour service plan: 2-10,000 hour and 1-30,000 hour maintenance | ||
Specifications | |||
| Front Panel Controls: | Power: On/Standby; Ventilation Modes: Control, Assist/Control, SIMV, CPAP; Breath Types: Volume Control, Pressure Control, Pressure Support, Spontaneous; Breath Rate; Tidal Volume; Inspiratory Time; PC/PS/Spontaneous Flow; Pressure Support; Sensitivity; Bias Flow; PEEP/CPAP; Control Lock: Easy or Hard unlock methods; Manual Breath: 1 x current settings | ||
| Power Requirements: | Electrical Requirements: 11 to 15 VDC; 90 to 250 VAC; 47 to 63 Hz; Pneumatic Requirements; Low Pressure Source: < 80 lpm Internal Battery: 60 mins. External Battery: 3 Hr or 9 Hr. Source Flow Oxygen (Medical Grade): 100 L/min | ||
| Languages: | English, French, German, Spanish, Italian, Swedish, Danish, Japanese | ||
| Alarms: | High Pressure Audible Alarm Delay: No Delay, 1 Breath, 2 Breaths Alarm Silence/Reset: 60 seconds; Ventilator Inoperative: Red LED; Disc/Sense; Low and Lost External Power; Low and Empty Internal Battery; High and Low Oxygen Inlet Pressure | ||
| Alarm Controls: | Apnea Interval: 10 to 60 seconds; High Pressure Limit: 5 to 100 cmH2O; Low Peak Pressure: Off, 1 to 60 cmH2O; Low Minute Volume: Off, 0.1 to 99 liters; Alarm Volume: 60 to 85 dBA at 1 meter | ||
| Non-Adjustable Alarms: | Disconnect; Low & Lost External Power; Low & Empty Internal Battery; High & Low Oxygen Inlet Pressure; Hardware Fault; Transducer Fault; Vent Reset; Setting Defaults | ||
| AC Power and Battery Indicators: | Red, Green, Yellow LEDs | ||
| Other Specifications: | Oxygen Source: Low pressure bleed in; Bias Flow: 10 lpm during exhalation; Over Pressure Relief: 110 cmH2O; Sub-Ambient Relief: Pressure Drop: -5 cmH2O; Control Lock: Easy or Hard Unlock methods; Manual Breath: single volume or pressure control breath NPPV Mode: Alarm package limited to: High Pressure, Disconnect, Defaults, Power Lost, Batter Low, Battery Empty, Apnea, Ventilator Inoperative Extended Features: NPPV Mode; Apnea Back-up Ventilation; Pressure Control and Pressure Support Variable Rise Time; Pressure Control and Pressure Support Variable Flow Termination Criteria; Pressure Support Variable Time Termination; Leak Compensation Other Monitored Variables: IE Ratio: 99:1 - 1:99; Calculated Peak Flow: 10 - 100 lpm; Patient Effort: Green LED; Monitor Scrolling | ||
Monitored Variables : | |||
| Pressure | Peak Inspiratory Pressure: 0 - 120 cmH2O, Mean Airway Pressure: 0 - 99 cmH2O, PEEP: 0 - 99 cmH2O, Airway Pressure Display: -10 to 108 cmH2O | ||
| Respiratory Rate | Total Breath Rate: 0 - 250 bpm | ||
| Tidal Volume | Exhalation Tidal Volume: 0 - 4000 ml | ||
| Minute Volume | Total Minute Volume: 0 - 99.9 liters | ||