| Puritan-Bennett Achieva | |||
Basic Description | |||
| Manufacturer: | Puritan Bennett Corporation | ||
| Ventilator Type: | Positive Pressure Invasive Noninvasive | ||
| Ventilator Class: | Intensive Care Unit (ICU) Subacute Home | ||
| Patient Population Range: | Pediatric Adult | ||
Description | |||
| Brief Description: | Portable volume ventilator with pressure support. | ||
Characteristics | |||
| Modes--Invasive: | General modes --- Assist/Control, SIMV, SPON | ||
| Operating Range of Controls: | Operating Altitude: Set from 0 to 4500 m; Volume: Adjustable from 50 to 2200 ml (1750 ml SIMV); Inspiratory Time: Set time from 0.2 to 5.0 seconds; Sensitivity: Set flow from 3 to 25 LPM. Set pressure from 1 to 15 cm H2O/hPa; Breath Rate: Set breathing rate: 1 to 80 BPM; Pressure: Set for 0 to 50 cm H2O/hPa; PEEP: Set for 0 and 3 to 20 cm H2O/hPa; Low Pressure: Set for 1 to 59 cm H2O/hPa; High Pressure: Set for 2 to 80 cm H2O/hPa; O2 Level: Set 21% to 100% | ||
| Size: | 10.75"H x 13.30"W x 15.60" D (27.3 cm x 33.8 cm x 39.6 cm) | ||
| Weight: | less than 32 lbs. | ||
| Other Characteristics: | Resistance Factor: Maximum of 4.5 cm H2O/hPa at 60 LPM; Emergency Pressure Relief: Approx. 90 ± 10 cm H2O/hPa; Flow: 2.0 LPM to 150 LPM (averaged) | ||
Characteristics (Non - Technical) | |||
| Other Comments: | Maintenance: Preventative maintenance by qualified professional every 6000 or recertification every 12 months, whichever comes first. | ||
Specifications | |||
| Front Panel Controls: | Alarm Silence/Reset; Battery Test; Standby; Ventilate; Menu/Esc; Start/Enter; Up and Down Arrow Keys; Patient Pressure Port; Patient Air Port 22 O.D./15mm I.D./Exhalation Valve Port | ||
| Power Requirements: | Power Line 100-240 VAC, 50/60 Hz, universal AC input; Power Usage Maximum: 75 W; Nominal: 10 W; External Battery: Approx. 20 hours operation under normal load; Internal Battery: At least 4 hours operation under normal load, backup use only I | ||
| Alarms: | Low Pressure/Apnea; High Pressure; Setting Error; Power Switchover; Low Power; O2 Failure | ||
| Non-Adjustable Alarms: | External Battery; Internal Battery; Battery Charge; Alarm Control; Assist/Spontaneous | ||
| Interface Ports: | Communication Port (RS232) | ||
| Other Specifications: | Operating Environment: 5C to 40C (41F to 104F), 10% to 90% RH non-condensing. Storage: -20C to 50C (-4F to 122F), 10% to 90% RH noncondensing. Allow at least 1 hour temperature stabilization before use when moving from operating to non-operating environment; the battery must be recharged every 30 days. | ||