VersaMed iVent201
   

Basic Description

 
 
Manufacturer:  VersaMed Inc.
Ventilator Type:  Positive Pressure
Invasive
Noninvasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Transport
Subacute
Home
CPAP
BiPAP
Patient Population Range:  Pediatric
Adult
Acute
 

Characteristics

 
Modes--Invasive:  Volume-controlled modes --- ASSIST/CONTROL; SIMV, PSV
Pressure-controlled modes --- ASSIST/CONTROL; SIMV, PSV
Supplemental modes --- CPAP, PSV, Adaptive Bilevel™ (noninvasive)
General modes --- Preset parameters by patient weight, Adaptive Flow™, Adaptive I-Time™, Apnea back-up, Sigh breath, Adjustable Rise Time, Easy Exhale™, 100% O2 suction
Operating Range of Controls:  Respiratory rate: 1-80BPM* Tidal volume: 50-2000ml* Inspiratory Pressure: 5-80cmH2O Inspiratory Time: Adaptive Time or 0.3-3sec Peak Flow: Adaptive Flow or 1-120L/min, spontaneous up to 180L/min Oxygen Mix (FiO2): 21%-100% PEEP: 0-40cmH2O Trigger Sensitivity: Flow Sens. Off or 1-20L/min, Pressure Sens. Off or -0.5-20cmH2O PSV: 0-60cmH2O Postitive Pressure Relief Valve: 80cmH2O *Available in some countries
Graphic Display:  MONITORING & DISPLAYED PARAMETERS Airway Pressure (Analog bar graph & numerical) Total Breath Rate Mean Airway Pressure I:E Ratio Exhaled Tidal Volume Peak Flow Inspiratory Time Electrical Power Source (external/internal) Battery Level Graphics and Diagnostics Software Package: Real Time Pressure, Flow and Volume Waveforms Waveform History Browse Trending of Monitored Data Pressure, Flow and Volume Loops Lung Mechanics (C, R)
Size:  Height: 33in/33cm Width: 9.5in/24cm Depth:10.3in/26cm Color Screen:8.4in diagonal
Weight:  22lb/10kg (with internal battery)
Other Characteristics:  iVent201 Ventilator Versions: iVent201 Ventilator Versions: IC+AB (Intensive Care + Facemask Ventilation in Adaptive Bi-Level): Pressure control (A/C and SIMV), Volume Control (A/C and SIMV), CPAP/PSV, Adaptive Bilevel, Pressure and Flow Waveforms, Trending and Respiratory Diagnostics, Internal Oxygen Blender with Sensor, High and Low Pressure Oxygen IC (Intensive Care): Pressure Control (A/C and SIMV), Volume Control (A/C and SIMV), CPAP/PSV, Pressure and Flow Waveforms, Trending and Respiratory Diagnostics, Internal Oxygen Blender, High and Low Pressure Oxygen AB (Adaptive Bi-Level): Adaptive Bi-Level, CPAP/PSV, Pressure and Flow Waveforms, Trending and Respiratory Diagnostics, Internal Oxygen Blender, High and Low Pressure Oxygen HC (Homecare): Volume Control (A/C and SIMV), CPAP/PSV, Pressure Waveform, Low Pressure Oxygen Indicators: Alarm Silence Icon and Timer Breath Type Icon Internal Battery Use Date and Time Hour Meter Battery Charge Level Need Calibration LED: On, Charge, Alarm Nebulizer port synchronized with inspiration Monitored Variables Pressure Analog bar graph and numerical PIP Total Respiratory Rate Exhaled Tidal Volume Exhaled Minute Volume I:E Ratio Delivered FiO2 Static pulmonary mechanics Built-in SpO2 monitoring Turbine technology, no need for air source (blended gas)

Characteristics (Non - Technical)

Approvals:  ASTM F1100-90, CSA C22.2 No. 601.1/601.2, IEC 60601-1, IEC 60601-1-2, EN 60601-1-12, EN 794-1/2/3, UL 2601.1
Other Comments:  Versamed is a CE marked company (mark 0473). Versamed markets its products under FDA regulations and as such has obtained market clearance under the following 510(k) allowances: K981668, K011957, and K042468. VersaMed iVent201 is FDA approved for ICU, Alternate Sites, Home, Emergency and Transport Use.

Specifications

Power Requirements:  External AC 100-240V, 50-60 Hz, Max 2.0A External DC 12-15V, Max 8.5A Inernal Battery: Sealed Lead-Acid 12V (7.8Ah) (rechargeable) Operating Time: up to 2 hours (depending on ventilation parameters)
Alarms:  Respiratory Rate (high/low) Minute Volume (high/low) Pressure (high/low) Low Vt (15%-85%) Apnea (0-120sec) FiO2 (high/low) Leak (0-100%)
Non-Adjustable Alarms:  Inverse I:E Ratio Low O2 Pressure AC Disconnect Low Battery Over Temperature Service Notice Patient Disconnect Check Sensor High PEEP
Interface Ports:  Remote Monitor (VGA) RS-232 Serial Port, 9 Pin DIN Keyboard Connector RJ11 Remote Alarm Connector
Other Specifications:  Oxygen (enrichment) Supply: High Pressure Supply: 40-60psi (2.8-4.2bar) Low Pressure: Max 80L/min or 0.5psi Environmental Specifications: Operating Temperature: 0-50C/32-120F Storage Temperature: -15 to 70C/-4 to 140F Relative Humidity: 15-95% at 30C/85F Water/Dust Resistance: IP54 (Splash Proof) Atmospheric Pressure: 430-825mmHg (15,000 feet) Vibration: IEC 68-2-6 and IEC 68-2-34, MIL-STD,810E Shock: IEC 68-2-27 (100g), MIL-STD-810E Total External Sound Level: 40-45dBa at one meter

Monitored Variables :

Pressure Analog bar graph and numerical PIP
Respiratory Rate Total rate
Tidal Volume Exhaled
Minute Volume Exhaled

Related Info 

Distributor |Educator | Manufacturer |Rental Agent | Service Center |
Parts Available | Product Markets | Service Centres |

Distributor:

Tri-anim
13170 Telfair Avenue
Sylmar, CA



Manufacturer:

VersaMed Inc.
2 Blue Hill Plaza
Pearl River, NY
USA