Puritan-Bennett Infant Star 950 Ventilator
   

Basic Description

 
 
Manufacturer:  Puritan Bennett Corporation
Ventilator Type:  Positive Pressure
Invasive
High Frequency
Ventilator Class:  Intensive Care Unit (ICU)
Subacute
CPAP
Patient Population Range:  Neonatal
Pediatric
 

Characteristics

 
Modes--Invasive:  Pressure-controlled modes --- ASSIST/CONTROL; SIMV; CPAP/BACKUP (these require Star Sync); IMV; CMV, CPAP
Flow-controlled modes --- CMV (optional); SIMV
Combined modes --- HFV, HFV + IMV
Waveforms:  Pressure: Rectangular; Flow: Rectangular
Operating Range of Controls:  Ventilator Rate: 1 to 150 bpm; PIP (Peak Inspiratory Pressure): 5 to 90 cmH2O; PEEP/CPAP: 0 to 24 cmH20; Inspiratory Time: 0.1 to 3.0 seconds; Flow Rate (Mandatory Breath): 4 to 40 Lpm; Background Flow: 2 to 32 Lpm (demand system provides up to 40 Lpm total flow; Proximal Airway Pressure: -10 to 120 cmH20; Mean Airway Pressure: 0 to 99.9 cmH2O; I:E Ratio: 1:0.1 to 1:99.9; Expiratory Time: 0.2 to 59.9; Duration Positive Pressure: Duration of positive pressure applied 1 cmH2O above baseline in seconds; High Frequency Rate: 2 to 22 Hz; Amplitude: Measured HFV pulse pressure 0 to 160 cmH2O
Size:  16.25" (41cm) H x 11.25" (29cm) W x 19.25" (49cm) D
Weight:  50 lb (23 kg)
Power: 

Characteristics (Non - Technical)

List Price:  $18785
Approvals:  CE (93/42/EEC) & (89/336/EEC), IEC 601-1-2 EMC, IEC 601-1, EN 60601-1, EN 60601-1-1, EN 60601-1-2, TUV, EN 29001 (ISO 9001), EN 46001, IEC 601-2-12
Other Comments:  Extended warranty/service agreement, high frequency conversion training seminars

Specifications

Front Panel Controls:  Ventilator Settings; Ventilator Rate; PIP (Peak Inspiratory Pressure); PEEP/CPAP; Inspiratory Time; Flow Rate (Mandatory Breath); Background Flow; Manual Breath; Proximal Airway Pressure:; Mean Airway Pressure; I:E Ratio; Expiratory Time; Duration Positive Pressure; High Frequency Rate: 2 to 22 Hz; Amplitude
Power Requirements:  Pneumatic Power: Air and/or Oxygen; 45 to 90 psig (241 to 621 kPa); Electrical Power: The Infant Star is continously powered by a 12-volt battery maintained fully charged when attached to AC power. During loss of AC power, the ventilator will continue to run for a minimum of 30 minutes. 90 to 105 VAC, 50/60 Hz, 100 watts; 105 to 127 VAC, 50/60 Hz, 100 watts; 198 to 264 VAC, 50/60 Hz, 100 watts; External 12 VDC Input; 11.8 V to 18 V @ 5 amps source; Battery Recharge Time; 1 hour (ventilator not in use); 1.5 hours (ventilator in use)
Alarms:  General System Failure; Additional General System Alarms; High Peak Pressure; Low Peak Pressure; Low PEEP; Apnea (optional)
Alarm Controls:  High Inspiratory Pressure: 5 to 105 cmH2O (maximum setting = PIP +15 cmH2O); Low Inspiratory Pressure: 3 to 60 cmH2O
Non-Adjustable Alarms:  Low PEEP/CPAP: Detects drop in PEEP/CPAP > 2 cmH2O; Airway Leak: Detects gross leaks in patient circuit; Obstructed Tube Alarms: AO1: HIP Alarm setting; AO2: Obstructed expiratory limb; 5 cmH2O above HIP Alarm setting; AO3: High expiratory resistance; <50% drop from PIP to PEEP within minimum expiratory time; AO4: High PEEP/CPAP; PEEP/CPAP +6 cmH2O for 5 seconds; AO5: Obstructed inspiratory limb or proximal line; 10 cmH2O about HIP Alarm setting; Insufficient Expiratory Time: Preset at 0.2 seconds or 0.3 seconds (determined by rate); Ventilator Inop + Error code: Malfunction detected by microprocessor diagnostics; Low O2 Pressure: Less than 45 psig (310 kPa) O2 pressure; Low Air Pressure: Less than 45 psig (310 kPa) air pressure; Low Battery: Imminent loss of battery power (5 min); External Power Loss: AC power loss - internal battery in use
Other Specifications:  30 minute battery duration. Alarms classified by urgency.

Monitored Variables :

Pressure During conventional ventilation airway pressure is displayed on an electronic analog gauge and during high frequency ventilation is displayed on an LED; peak inspiratory pressure, PEEP/CPAP, mean airway pressure, duration of positive pressure are all displayed on LEDs

Related Info 

Distributor |Educator | Manufacturer |Rental Agent | Service Center |
Parts Available | Product Markets | Service Centres |

Distributor:

Huff and Puff Victoria Inc.
PO Box 168
Melbourne,
Australia



Distributor:

Pioneer Medical, Inc.
566 Mainstream Drive
Nashville, TN



Manufacturer:

Puritan Bennett Corporation
2200 Faraday Avenue
Carlsbad, CA
USA