| Respironics Esprit Ventilator | |||
Basic Description | |||
| Manufacturer: | Respironics | ||
| Ventilator Type: | Positive Pressure Invasive Noninvasive | ||
| Ventilator Class: | Intensive Care Unit (ICU) Transport Subacute CPAP | ||
| Patient Population Range: | Pediatric Adult Acute | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- ASSIST/CONTROL; SIMV (PSV) Pressure-controlled modes --- ASSIST/CONTROL; SIMV + PSV; CPAP | ||
| Operating Range of Controls: | Tidal Volume 50 ml to 2500 ml; Respiratory Rate 1 to 80 BPM; Plateau 0 to 2.0 sec.; Peak Inspiratory Flow 3 to 140 LPM (Tubing compliance compensated); Spontaneous Flow Up to 200 LPM (Uncompensated); PEEP/CPAP 0 to 35 cm H2O; PSV Pressure 0 to 100 cm H2O; Risetime 0.1 to 0.9 sec.; Inspiratory Time 0.1 to 9.9 sec; | ||
| Triggering: | Inspiratory --- Pressure Sensitivity -20 to -0.1 cm H2O; Flow Sensitivity 0.5 to 20 LPM; Base Flow 3 LPM above flow sensitivity (During flow triggering); Expiratory --- Exhalation Sensitivity 10 to 45% of peak flow ; Flow Wave Forms Descending; Percent O2 21 to 100%; Patient Type Adult, Pediatric; Apnea Backup Rate 1 to 80 BPM; | ||
| Graphic Display: | Peak Inspiratory Pressure: -20 to 130 cm H2O; Mean Airway Pressure: -20 to 120 cm H20; Plateau Pressure: -20 to 130 cm H2O; End Expiratory Pressure: -20 to 99.9 cm H2O; End Expiratory Pressure: -20 to 120 cm H2O; Expiratory Pause Pressure: -20 to 120 cm H2O; | ||
| Size: | 40.5 x 35.6 x 61.0 (cm) | ||
| Weight: | 30 (kg); With Backup Battery 39 (kg) | ||
Characteristics (Non - Technical) | |||
| Warranty Period: | 2 years | ||
Specifications | |||
| Front Panel Controls: | Alarm Reset; 100% O2; Expiratory Hold; Accept; Screen Lock; Alarm Silence; Manual Inspiration; Display Brightness; Value Knob; Alarm Volume Control(Min to max); | ||
| Alarms: | High Inspiratory Pressure: 10 to 105 cm H2O; Low Inspiratory Pressure: 3 to 105 cm H2O; Low PEEP/CPAP Pressure: 0 to 35 cm H2O; High Respiratory Rate: 0 to 150 BPM; Low Exhaled Mandatory Tidal Volume: 0 ml to 2500 ml; Low Exhaled Spontaneous Tidal Volume: 0 ml to 2500 ml; Low Exhaled Minute Volume: 0 to 60 LPM; High Exhaled Minute Volume: 0 to 60 LPM; Apnea Interval: 10 to 60 sec. | ||
| Alarm Status Indicators: | Normal Indicator; Safety Valve Open; Ventilator Inoperative; Screen Locked; Alarm Silence; High Urgency Alarm; | ||
| Non-Adjustable Alarms: | Circuit Occlusion; Low O2 Inlet Pressure; High Delivered O2%; Air Source Inoperative; Inspiratory Time Too Long; Low Delivered O2%; | ||
| AC Power and Battery Indicators: | Loss of Power (Audible); Back-up Battery: Charging, In Use, Low; Low of AC Power (Audible); External Battery In Use; Mains; | ||
| Interface Ports: | RS-232 Output and Nurse Call Outpu; | ||
| Mounting Options: | Deluxe or Standard Cart; Wall Mount; Table Top; | ||
| Other Specifications: | Software Update/Upgrade --- Via Serial Interface (Field updateable / upgradeable); | ||
Monitored Variables : | |||
| Respiratory Rate | 0 to 150 BPM | ||
| Tidal Volume | 0 to 9999 ml | ||
| Minute Volume | 0 to 99.9 LPM | ||
Related Info | |||