Respironics Esprit Ventilator
   

Basic Description

 
 
Manufacturer:  Respironics
Ventilator Type:  Positive Pressure
Invasive
Noninvasive
Ventilator Class:  Intensive Care Unit (ICU)
Transport
Subacute
CPAP
Patient Population Range:  Pediatric
Adult
Acute
 

Characteristics

 
Modes--Invasive:  Volume-controlled modes --- ASSIST/CONTROL; SIMV (PSV)
Pressure-controlled modes --- ASSIST/CONTROL; SIMV + PSV; CPAP
Operating Range of Controls:  Tidal Volume 50 ml to 2500 ml; Respiratory Rate 1 to 80 BPM; Plateau 0 to 2.0 sec.; Peak Inspiratory Flow 3 to 140 LPM (Tubing compliance compensated); Spontaneous Flow Up to 200 LPM (Uncompensated); PEEP/CPAP 0 to 35 cm H2O; PSV Pressure 0 to 100 cm H2O; Risetime 0.1 to 0.9 sec.; Inspiratory Time 0.1 to 9.9 sec;
Triggering:  Inspiratory --- Pressure Sensitivity -20 to -0.1 cm H2O; Flow Sensitivity 0.5 to 20 LPM; Base Flow 3 LPM above flow sensitivity (During flow triggering); Expiratory --- Exhalation Sensitivity 10 to 45% of peak flow ; Flow Wave Forms Descending; Percent O2 21 to 100%; Patient Type Adult, Pediatric; Apnea Backup Rate 1 to 80 BPM;
Graphic Display:  Peak Inspiratory Pressure: -20 to 130 cm H2O; Mean Airway Pressure: -20 to 120 cm H20; Plateau Pressure: -20 to 130 cm H2O; End Expiratory Pressure: -20 to 99.9 cm H2O; End Expiratory Pressure: -20 to 120 cm H2O; Expiratory Pause Pressure: -20 to 120 cm H2O;
Size:  40.5 x 35.6 x 61.0 (cm)
Weight:  30 (kg); With Backup Battery 39 (kg)

Characteristics (Non - Technical)

Warranty Period:  2 years

Specifications

Front Panel Controls:  Alarm Reset; 100% O2; Expiratory Hold; Accept; Screen Lock; Alarm Silence; Manual Inspiration; Display Brightness; Value Knob; Alarm Volume Control(Min to max);
Alarms:  High Inspiratory Pressure: 10 to 105 cm H2O; Low Inspiratory Pressure: 3 to 105 cm H2O; Low PEEP/CPAP Pressure: 0 to 35 cm H2O; High Respiratory Rate: 0 to 150 BPM; Low Exhaled Mandatory Tidal Volume: 0 ml to 2500 ml; Low Exhaled Spontaneous Tidal Volume: 0 ml to 2500 ml; Low Exhaled Minute Volume: 0 to 60 LPM; High Exhaled Minute Volume: 0 to 60 LPM; Apnea Interval: 10 to 60 sec.
Alarm Status Indicators:  Normal Indicator; Safety Valve Open; Ventilator Inoperative; Screen Locked; Alarm Silence; High Urgency Alarm;
Non-Adjustable Alarms:  Circuit Occlusion; Low O2 Inlet Pressure; High Delivered O2%; Air Source Inoperative; Inspiratory Time Too Long; Low Delivered O2%;
AC Power and Battery Indicators:  Loss of Power (Audible); Back-up Battery: Charging, In Use, Low; Low of AC Power (Audible); External Battery In Use; Mains;
Interface Ports:  RS-232 Output and Nurse Call Outpu;
Mounting Options:  Deluxe or Standard Cart; Wall Mount; Table Top;
Other Specifications:  Software Update/Upgrade --- Via Serial Interface (Field updateable / upgradeable);

Monitored Variables :

Respiratory Rate 0 to 150 BPM
Tidal Volume 0 to 9999 ml
Minute Volume 0 to 99.9 LPM

Related Info 

Distributor |Educator | Manufacturer |Rental Agent | Service Center |
Parts Available | Product Markets | Service Centres |

Distributor:

Pioneer Medical, Inc.
566 Mainstream Drive
Nashville, TN



Manufacturer:

Respironics
1501 Ardmore Blvd.
Pittsburgh, PA