| Sechrist 500A Hyperbaric Ventilator | |||
Basic Description | |||
| Manufacturer: | Sechrist Industries | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Subacute | ||
| Patient Population Range: | Adult | ||
Description | |||
| Brief Description: | The Sechrist 500A Hyperbaric Ventilator is designed for use with the Sechrist Monoplace Hyperbaric Chamber, and automatically compensates for pressure changes in the chamber by automatically adjusting delivery pressure to maintain tidal volumes without affecting inspiratory and expiratory times. | ||
Characteristics | |||
| Operating Range of Controls: | Minute Volume Range:o 0-15 LPM at 3 A.T.A. (2.1 kg/cm2); Tidal Volume Range: 0-1.5 Liters at 3 A.T.A. (2.1 kg/cm2); Frequency Range: 8-30 breaths per minute; Inspiratory Time Range: 1.0-3.5 seconds; Expiratory Time Range: 1.0-5.0 seconds; I:E Ratio Range: 1:5-3.5:1; Inspiratory Flow Range: 0-100 LPM at ambient; 0-60 LPM at 3 A.T.A. (2.1 kg/cm2); Inspiratory Pressure Limit: 20 - 80 cm H2O adjustable Safety Pressure Relief Valve | ||
| Other Characteristics: | Manual Control: Manual push button for inspiration. Inspiration occurs as long as the button is pushed. Amount of depression controls flow. Bypasses all internal ventilator circuitry. Airway Pressure Gauge: 0-80 cm H2O Manometer | ||
Characteristics (Non - Technical) | |||
| Recommended Humid: | 500 ml nebulizer | ||
Specifications | |||
| Power Requirements: | Power Source: 70-80 p.s.i. (4.9-5.6 kg/cm2 or 482.7-551.6 kPa) Medical Breathing Oxygen Power Consumption: 26 LPM (max) during inspiratory phase | ||
| Other Specifications: | Manual Control: Manual push button for inspiration. Inspiration occurs as long as the button is pushed. Amount of depression controls flow. Bypasses all internal ventilator circuitry. Humidifier: 500 ml nebulizer; Airway Pressure Gauge: 0-80 cm H2O Manometer | ||
Related Info | |||