Siemens Servo 300A Ventilator with Automode
   

Basic Description

 
 
Manufacturer:  MAQUET GmbH & Co. KG
Ventilator Type:  Positive Pressure
Invasive
Conventional
High Frequency
Ventilator Class:  Intensive Care Unit (ICU)
Patient Population Range:  Neonatal
Pediatric
Adult
 

Description

 
Features:  Automode feature enhances the process of weaning
Brief Description:  Go Beyond The Static Specs! Click Here For Special Online Interactive Content! (Requires Flash Player 5.0)
 

Characteristics

 
Modes--Invasive:  Volume-controlled modes --- SIMV, SIMV + Pressure Support
Pressure-controlled modes --- SIMV, SIMV + Pressure Support
Combined modes --- Automatic mode switch between Pressure Control and Pressure Support; Volume Control and Volume Support; Pressure Regulated Volume Control and Volume Support
Operating Range of Controls:  CMV frequency: 5 - 150 breaths/min; SMV frequency: 0.5 - 40 breaths/min; Inspiration time: 10% - 80% of breath cycle time (stepless); Pause time: 0% - 30% of breath cycle time (stepless); Pressure control: 0 - 100 cm H2O; Pressure support: 0 - 100 cm H2O; PEEP: 0 - 50 cm H2O; Trigger sensitivity: Flow: 3 - 32 ml/s (green marked area); Pressure: -17 - 0 cm H2O; Trigger bias flow: Neonate: 8 ml/s (0.5 l/min); Pediatric: 16 ml/s (1 l/min); Adult: 32 ml/s (2 l/min); Inspiration rise time: 0% - 10% of breath cycle time (stepless); Preset Tidal Volume: 2 - 4000 ml; Preset Minute Volume: 0.2 - 60 l/min; Oxygen breaths: 100% for 20 breaths or max 1 min; Oxygen concentration: 21% to 100%
Triggering:  Flow, pressure
Size:  Patient Unit: 242mm W x 370mm D x 240mm H; Control Unit: 431mm W x 150mm D x 325mm H
Weight:  approximately 24 kg
Other Characteristics:  The switch from controlled to supported mode is caused by patient triggering efforts; The switch from supported to controlled mode happens after 12 seconds apnea (Adult), 5 seconds apnea (Pediatric), or 5 seconds apnea (Neonate) Other monitored variables; Inspiratory period (I:E ratio): set; O2 ratio: set, measured: Gas supply pressure: measured (air and O2); Battery voltage: measured (internal and external)

Characteristics (Non - Technical)

Approvals:  Class 1 equipment Type B according to IEC 601-1/EN 60 601-1

Specifications

Front Panel Controls:  CMV frequency; SMV frequency; Inspiration time; Pause time; Pressure control; Pressure support; PEEP; Trigger sensitivity; Trigger bias flow; Inspiration rise time; Preset Tidal Volume; Preset Minute Volume; Oxygen breaths; Start breath: Initiation of 1 breath in all modes; Pause hold: Inspiration or expiration; Patient range: Neonate/Pediatric/Adult; Oxygen concentration: Automode: On/Off
Power Requirements:  Inlet gas pressure: 2 - 6.5 bar (29 - 94 PSI), air and O2; Power supply: 100, 120, 220, and 240 VAC +/- 10%; 50 - 60 Hz; Battery: Back-up time: Approximately 30 minutes; Recharge time: Approximately 10 hours; External battery input 24 V DC; Power consumption: 50 W (stand alone)
Alarms:  Alarm silence; Alarm reset
Alarm Controls:  Airway pressure (upper): 15 - 120 cm H2O; O2 concentration (high and low): set value +/- 6%; Expired minute volume: Upper alarm limit: Adult/pediatric: 0 - 60 l/min; Neonate: 0 - 6 l/min; Lower alarm limit: Adult/pediatric: 0.3 - 40 l/min; Neonate: 0.06 - 4 l/min
Non-Adjustable Alarms:  High continuous pressure: Apnea: Neonate: 10 seconds; Pediatric: 15 seconds; Adult: 20 seconds; Gas supply; Battery; Technical
Interface Ports:  Serial port: RS232 communication; Analog terminal: For analog outputs; Master/slave connection: For ILV (Independent Lung Ventilation) synchornization; Auxiliary equipment: for optional equipment
Other Specifications:  Flow range: 0.1 ml/s - 3 l/s; Operating Conditions: Temperature range: +15°C to +35°C (60°F to 90°F); Relative humidity: 30% to 50%; Atmospheric pressure: 700 to 1060 hPa (525 to 705mm Hg); Non-Operating Conditions: Impact: In accordance with IEC 68-2-29Eb; Peak acceleration: 15 g; Pulse duration: 6 ms; Number of impacts: 1000; Storage temperature range: -25°C to +70°C (-13°F to 158°F)

Monitored Variables :

Pressure Peak, Mean, Pause, End expiratory, Airway
Respiratory Rate Set, Measured
Tidal Volume Set, Inspiratory, Expiratory
Minute Volume Expiratory

Related Info 

Distributor |Educator | Manufacturer |Rental Agent |Service Center |
Parts Available | Product Markets | Service Centres |

Distributor:

Freedom Medical
208 Carter Drive
West Chester, PA



Distributor:

Pioneer Medical, Inc.
566 Mainstream Drive
Nashville, TN



Manufacturer:

MAQUET GmbH & Co. KG
Kehlerstraße 31
Rastatt,
Germany



Rental Agent:

Freedom Medical
208 Carter Drive
West Chester, PA