| Siemens Servo 300A Ventilator with Automode | |||
Basic Description | |||
| Manufacturer: | MAQUET GmbH & Co. KG | ||
| Ventilator Type: | Positive Pressure Invasive Conventional High Frequency | ||
| Ventilator Class: | Intensive Care Unit (ICU) | ||
| Patient Population Range: | Neonatal Pediatric Adult | ||
Description | |||
| Features: | Automode feature enhances the process of weaning | ||
| Brief Description: | Go Beyond The Static Specs! Click Here For Special Online Interactive Content! (Requires Flash Player 5.0) | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- SIMV, SIMV + Pressure Support Pressure-controlled modes --- SIMV, SIMV + Pressure Support Combined modes --- Automatic mode switch between Pressure Control and Pressure Support; Volume Control and Volume Support; Pressure Regulated Volume Control and Volume Support | ||
| Operating Range of Controls: | CMV frequency: 5 - 150 breaths/min; SMV frequency: 0.5 - 40 breaths/min; Inspiration time: 10% - 80% of breath cycle time (stepless); Pause time: 0% - 30% of breath cycle time (stepless); Pressure control: 0 - 100 cm H2O; Pressure support: 0 - 100 cm H2O; PEEP: 0 - 50 cm H2O; Trigger sensitivity: Flow: 3 - 32 ml/s (green marked area); Pressure: -17 - 0 cm H2O; Trigger bias flow: Neonate: 8 ml/s (0.5 l/min); Pediatric: 16 ml/s (1 l/min); Adult: 32 ml/s (2 l/min); Inspiration rise time: 0% - 10% of breath cycle time (stepless); Preset Tidal Volume: 2 - 4000 ml; Preset Minute Volume: 0.2 - 60 l/min; Oxygen breaths: 100% for 20 breaths or max 1 min; Oxygen concentration: 21% to 100% | ||
| Triggering: | Flow, pressure | ||
| Size: | Patient Unit: 242mm W x 370mm D x 240mm H; Control Unit: 431mm W x 150mm D x 325mm H | ||
| Weight: | approximately 24 kg | ||
| Other Characteristics: | The switch from controlled to supported mode is caused by patient triggering efforts; The switch from supported to controlled mode happens after 12 seconds apnea (Adult), 5 seconds apnea (Pediatric), or 5 seconds apnea (Neonate) Other monitored variables; Inspiratory period (I:E ratio): set; O2 ratio: set, measured: Gas supply pressure: measured (air and O2); Battery voltage: measured (internal and external) | ||
Characteristics (Non - Technical) | |||
| Approvals: | Class 1 equipment Type B according to IEC 601-1/EN 60 601-1 | ||
Specifications | |||
| Front Panel Controls: | CMV frequency; SMV frequency; Inspiration time; Pause time; Pressure control; Pressure support; PEEP; Trigger sensitivity; Trigger bias flow; Inspiration rise time; Preset Tidal Volume; Preset Minute Volume; Oxygen breaths; Start breath: Initiation of 1 breath in all modes; Pause hold: Inspiration or expiration; Patient range: Neonate/Pediatric/Adult; Oxygen concentration: Automode: On/Off | ||
| Power Requirements: | Inlet gas pressure: 2 - 6.5 bar (29 - 94 PSI), air and O2; Power supply: 100, 120, 220, and 240 VAC +/- 10%; 50 - 60 Hz; Battery: Back-up time: Approximately 30 minutes; Recharge time: Approximately 10 hours; External battery input 24 V DC; Power consumption: 50 W (stand alone) | ||
| Alarms: | Alarm silence; Alarm reset | ||
| Alarm Controls: | Airway pressure (upper): 15 - 120 cm H2O; O2 concentration (high and low): set value +/- 6%; Expired minute volume: Upper alarm limit: Adult/pediatric: 0 - 60 l/min; Neonate: 0 - 6 l/min; Lower alarm limit: Adult/pediatric: 0.3 - 40 l/min; Neonate: 0.06 - 4 l/min | ||
| Non-Adjustable Alarms: | High continuous pressure: Apnea: Neonate: 10 seconds; Pediatric: 15 seconds; Adult: 20 seconds; Gas supply; Battery; Technical | ||
| Interface Ports: | Serial port: RS232 communication; Analog terminal: For analog outputs; Master/slave connection: For ILV (Independent Lung Ventilation) synchornization; Auxiliary equipment: for optional equipment | ||
| Other Specifications: | Flow range: 0.1 ml/s - 3 l/s; Operating Conditions: Temperature range: +15°C to +35°C (60°F to 90°F); Relative humidity: 30% to 50%; Atmospheric pressure: 700 to 1060 hPa (525 to 705mm Hg); Non-Operating Conditions: Impact: In accordance with IEC 68-2-29Eb; Peak acceleration: 15 g; Pulse duration: 6 ms; Number of impacts: 1000; Storage temperature range: -25°C to +70°C (-13°F to 158°F) | ||
Monitored Variables : | |||
| Pressure | Peak, Mean, Pause, End expiratory, Airway | ||
| Respiratory Rate | Set, Measured | ||
| Tidal Volume | Set, Inspiratory, Expiratory | ||
| Minute Volume | Expiratory | ||