| Newport Wave E200 Ventilator | |||
Basic Description | |||
| Manufacturer: | Newport Medical Instruments | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) Transport Subacute | ||
| Patient Population Range: | Neonatal Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- ASSIST/CONTROL; SIMV (PSV) Pressure-controlled modes --- ASSIST/CONTROL; SIMV (PSV); CPAP | ||
| Waveforms: | Rectangular flow; | ||
| Operating Range of Controls: | %O2: 21 to 100% Mechanical Flow: 1 to 100 L/min; Inspiratory Time: 0.1 to 3.0 sec; Inspiratory Pause: 1, 10, 20 & 30% of I.T.; Rate: 1 to 100 b/min; Tidal Volume: .01 to 2.0L; Pressure Control Level: 0 to 75 cmH2O; Pressure Support Level: 0 to 60 cmH2O; PEEP/CPAP: 0 to 45 cmH2O; Bias Flow: 0 to 30 L/min; Spont/Demand Flow: 0 to 160 L/min; Sensitivity: 0 to -5 cmH2O; Max Inverse I:E: 3:1; Manometer Range: -10 to 120 cmH2O; Pneumatic Pressure Relief: 0 to 120 cmH2O; | ||
| Graphic Display: | Visual alarm LED's; Electronic pressure gauge; | ||
| Size: | 30 x 35 x 38 cm | ||
| Weight: | 25 kg | ||
Characteristics (Non - Technical) | |||
| Other Comments: | Preventative maintenance every 3,000 hours or once/year, overhaul every 15,000 hours or every 5 years. | ||
Specifications | |||
| Power Requirements: | 100, 120, 220, 240 VAC / 47 to 63 Hz, 125 VA MAX | ||
| Alarms: | Input power; Inspiratory time too long; Ventalator inoperative; | ||
| Alarm Controls: | Low pressure: 0 to 110cmH2O; High pressure: 5 to 120 cmH2O; High minute volume: 1 to 50 L/min; Low minutevolume: 0 to 49 L/min; | ||
Monitored Variables : | |||
| Pressure | |||
| Respiratory Rate | |||
| Tidal Volume | |||
| Minute Volume | |||