| Newport Breeze E150 Ventilator | |||
Basic Description | |||
| Manufacturer: | Newport Medical Instruments | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) Transport Subacute CPAP | ||
| Patient Population Range: | Neonatal Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- ASSIST/CONTROL; SIMV Pressure-controlled modes --- ASSIST/CONTROL; SIMV; CPAP | ||
| Operating Range of Controls: | %O2: 21 to 100%; Flow: 1 to 120 L/min; Inspiratory Time: 0.1 to 3.0 sec; I:E Ratio: 1:99 to 4:1; Set Rate: 1 to 150 b/min; Tidal Volume: 0.01 to 2.0 L; PIP(PCV): 0 to 60 cmH2O; PEEP/CPAP: 0 to 60 cmH2O; Spontaneous Flow: 0 to +40 L/min; Trigger Level: -10 to +60 cmH2O; Pneumatic Pressure Relief: 0 to 120 cmH2O; | ||
| Size: | 33 x 28 x 25 (cm) | ||
| Weight: | 13.6 (kg) | ||
| Power: | |||
Characteristics (Non - Technical) | |||
| Approvals: | CE, NSAI, ISO | ||
| Other Comments: | Preventative maintenance every 3,000 hours or once/year, overhaul every 15,000 hours or every 5 years. | ||
Specifications | |||
| Power Requirements: | A.C. Power: 100, 120, 220, or 240 VAC +- 10%, 47 to 63 Hz 45 watts max; | ||
| Alarms: | High/Low Pressure: Audible and Visual; Apnea: Audible and Visual; Low CPAP: Audible and Visual; Low Battery: Audible and Visual; Gas Supply Source Failure: Audible; Power Failure: Audible; Power Down: Audible | ||
| Other Specifications: | Internal Battery--1 hour minimum, recharge 18 hours maximum | ||
Monitored Variables : | |||
| Pressure | |||
| Respiratory Rate | |||