Dräger Babylog 8000 plus Ventilator
   

Basic Description

 
 
Manufacturer:  Dräger Medical, Inc.
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
CPAP
Patient Population Range:  Neonatal
 

Characteristics

 
Modes--Invasive:  Volume-controlled modes --- ASSIST/CONTROL; SIMV; MMV
Pressure-controlled modes --- ASSIST/CONTROL; SIMV; CPAP (0-25 cmH20)
General modes --- Pressure Support, Volume Guarantee (maintains pre-set volume and pressure simultaneously), Leak Adapted Pressure Support
Operating Range of Controls:  Inspiratory Pressure: 10-80 cmH2O; PEEP: 0-25 cmH2O; Inspiratory Time: 0.1-2.0 sec.; Expiratory Time: 0.7-30 sec.; 02 concentration: 21%-100%; Inspiratory Flow: 1-30 LPM; Expiratory Flow: 1-30 LPM; Tidal Volume (set); 2-100 LPM
Triggering:  Spontaneous breathing is detected with a flow measurement. Trigger sensitivity can be set from 1-10. At maximum sensitivity, V trigger is 0 and inspiratory flow will trigger a breath when it reaches 0.2 LPM. Trigger response time is 40-60 msec.
Graphic Display:  P and V waveform and trends;
Size:  8" x 11" x 15.4"

Characteristics (Non - Technical)

Other Comments:  Standard preventative maintenance for parts: less than $300/5 yr. W/internal DC option less than $800/5 yr

Specifications

Front Panel Controls:  Operating Modes; O2 concentration; Inspiratory Time; Inspiratory Flow; Peak Inspiratory Pressure; PEEP/CPAP
Other Specifications:  Alarms classified by urgency

Monitored Variables :

Pressure
Respiratory Rate
Tidal Volume
Minute Volume

Related Info 