| Dräger Babylog 8000 plus Ventilator | |||
Basic Description | |||
| Manufacturer: | Dräger Medical, Inc. | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) CPAP | ||
| Patient Population Range: | Neonatal | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- ASSIST/CONTROL; SIMV; MMV Pressure-controlled modes --- ASSIST/CONTROL; SIMV; CPAP (0-25 cmH20) General modes --- Pressure Support, Volume Guarantee (maintains pre-set volume and pressure simultaneously), Leak Adapted Pressure Support | ||
| Operating Range of Controls: | Inspiratory Pressure: 10-80 cmH2O; PEEP: 0-25 cmH2O; Inspiratory Time: 0.1-2.0 sec.; Expiratory Time: 0.7-30 sec.; 02 concentration: 21%-100%; Inspiratory Flow: 1-30 LPM; Expiratory Flow: 1-30 LPM; Tidal Volume (set); 2-100 LPM | ||
| Triggering: | Spontaneous breathing is detected with a flow measurement. Trigger sensitivity can be set from 1-10. At maximum sensitivity, V trigger is 0 and inspiratory flow will trigger a breath when it reaches 0.2 LPM. Trigger response time is 40-60 msec. | ||
| Graphic Display: | P and V waveform and trends; | ||
| Size: | 8" x 11" x 15.4" | ||
Characteristics (Non - Technical) | |||
| Other Comments: | Standard preventative maintenance for parts: less than $300/5 yr. W/internal DC option less than $800/5 yr | ||
Specifications | |||
| Front Panel Controls: | Operating Modes; O2 concentration; Inspiratory Time; Inspiratory Flow; Peak Inspiratory Pressure; PEEP/CPAP | ||
| Other Specifications: | Alarms classified by urgency | ||
Monitored Variables : | |||
| Pressure | |||
| Respiratory Rate | |||
| Tidal Volume | |||
| Minute Volume | |||
Related Info | |||