| Bird TBird * Legacy Ventilator | |||
Basic Description | |||
| Manufacturer: | Bird Products Corporation | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Subacute Home CPAP | ||
| Patient Population Range: | Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- PSV Pressure-controlled modes --- ASSIST/CONTROL; SIMV (PSV); CPAP | ||
| Operating Range of Controls: | Tidal Volume: 50 to 2000 ml; Breath Rate: 2 to 80 bpm; Peak Flow: 10 to 140 lpm; Spontaneous Peak Flow: 180 lpm max.; Sensitivity: 1 to 8 lpm, OFF; PEEP/CPAP: 0 to 30 cmH2O; Pressure Support: Off, 1 to 60 cmH2O; Oxygen Percent (VSO2 only): 21 to 100%; Bias Flow: 10 to 20 lpm; Sigh: On / Off, 1 sigh breath every 100 breaths or 7 minutes, whichever occurs first, 1.5 x Set Tidal Volume; Manual Breath: x1; 100% O2 (O2 only): 21 to 100%; Flow Cal: On / Off; Pressure Relief: 20 to 130 cmH2O | ||
| Triggering: | Flow | ||
| Size: | 32 W x 35.6 D x 33 H cm (12.6" W x 14" D x13" H) | ||
| Weight: | 19.1 kg (42 lbs.) | ||
| Other Characteristics: | Monitors I:E Ratio: 99:1 to 1:99 | ||
Characteristics (Non - Technical) | |||
| List Price: | $8900 | ||
| Approvals: | IEC 601-1, IEC 601-1-2, IEC 601-2-12, UL 2601-1 | ||
| Other Comments: | Equipment total: $10550; * TBird is a registered trademark in the US and some other countries | ||
Specifications | |||
| Front Panel Controls: | Power: On/Standby; Modes: Control, Assist/Control, SIMV, CPAP; Flow Pattern: Decelerating; Tidal Volume; Breath Rate; Peak Flow; Sensitivity; PEEP/CPAP; Pressure Support; Oxygen Percent (VSO2; Bias Flow; Sigh; Manual Breath; 100% O2; Flow Cal; Pressure Relief | ||
| Power Requirements: | Electrical Specifications; Input Voltage Requirements: AC: Voltage: 100-240 VAC; Frequency: 47-63 Hz; Power: 200 watts; Power Factor: 97; Input Fuse Specification Information: 100-250V~: Quantity: 2; Current Rating: 3-15 Amperes; Voltage Rating: 250 Volts; Type: 5 x 20 mm, IEC 127 Sheet 1 (Type F, fast acting, high-interrupting capacity); Pneumatic Specifications: Oxygen Inlet: Legacy Oxygen Bleed in: 0-80 lpm; Legacy O2 Pressurized Oxygen: 40-60 PSI (2,8 to 4,2 Bar) | ||
| Alarms: | Alarm Silence: 60 second max.; Alarm Volume: 65 - 85 dba; <50 dBa (rms) at one meter High pressure alarm triggers the release of circuit pressure | ||
| Alarm Controls: | High Pressure Limit: 5 to 120 cmH2O; Low Pressure Limit :Off, 2 to 60 cmH2O; Low Minute Volume: 0.0 to 99.9 L; Apnea Interval: 10 to 60 seconds ; Backup Breath Rate: 12 bpm or set breath rate, whichever is greater | ||
| Non-Adjustable Alarms: | Low O2 (O2 only): 35 PSIG (2.4 Bar); High O2 (O2 only): 65 PSIG (4.3 Bar); Ventilator Inoperative: Red Indicator | ||
| AC Power and Battery Indicators: | Displays the current power source and the status of the internal battery | ||
| Mounting Options: | Can be mounted behind most wheelchairs and scooters | ||
| Other Specifications: | Environmental Temperature: Operating: +5 to 40° C; Storage: -20 to 60° C; Humidity: Operating: 15% to 95% relative, noncondensing; Storage: 10% to 95% relative, noncondensing | ||
Monitored Variables : | |||
| Pressure | Airway Pressure Manometer: -20 to 120 cmH2O; Peak Inspiratory Pressure (PIP): 0 to 140 cmH2O; Mean Airway Pressure (MAP): 0 to 99 cmH2O; Positive End Expiratory Pressure (PEEP); cmH2O: 0 to 99 cmH2O | ||
| Respiratory Rate | Total Breath Rate (f): 0 to 250 bpm | ||
| Tidal Volume | 0 to 4000 ml | ||
| Minute Volume | 0 to 99.9 L | ||