| Bird TBird VS/TBird VSO2 Ventilator System | |||
Basic Description | |||
| Manufacturer: | Bird Products Corporation | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) Transport Subacute Home CPAP | ||
| Patient Population Range: | Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- CONTROL; ASSIST/CONTROL; SIMV (PSV); CPAP; MMV Pressure-controlled modes --- ASSIST/CONTROL; SIMV (PSV) | ||
| Operating Range of Controls: | Tidal Volume: 50 to 2000 ml; Breath Rate: 2 to 80 bpm; Peak Flow: 10 to 40 lpm; Spontaneous Peak Flow: 180 lpm max.; Sensitivity: 1 to 8 lpm, OFF; PEEP/CPAP: 0 to 30 cmH2O; Pressure Support: Off, 1-60 cmH2O; Oxygen Percent (VSO2 only): 21-100%; Bias Flow: 10 to 20 lpm; Sigh: On / Off, 1 sigh breath every 100 breaths or 7 minutes, whichever occurs first, 1.5 x Set Tidal Volume; Manual Breath: x1; Remote Alarm (optional): On / Off; Inspiratory Hold: 6 second max.; 100% O2 (VSO2 only): On/ Off, 3 minute max.; Flow Cal: On / Off; Pressure Relief: 20-130 cmH2O | ||
| Size: | 33.0 W x 36.8 D x 27.9 H cm (13" W x 14.5" D x 11" H ) | ||
| Weight: | 14 kg (31 lbs.) | ||
| Other Characteristics: | Operating Environment: 5°C to 40°C (41° F to 104°F), 10% to 90% RH noncondensing Storage Environment: -20°C to 50°C (-4°F to 122°F), 10% to 90% RH noncondensing; allow at least 1 hour temperature stabilization before use when moving from non-operating to operating environment; the battery must be recharged every 30 days | ||
Specifications | |||
| Front Panel Controls: | Front Panel Controls/Connections: Alarm Silence/Reset; Battery Test; Standby; Ventilate; Menu/Esc; Start/Enter; Up and Down Arrow Keys; Patient Pressure Port; Patient Air Port: 22mm O.D./15 mm I.D.; Exhalation Valve Port | ||
| Alarms: | High Pressure Limit: 5 to 120 cmH2O; Low Pressure Limit: Off, 2 to 60 cmH2O; Low Minute Volume: 0.1 to 99.9 L; Apnea Interval: 10 to 60 seconds; Backup Breath Rate: 12 bpm or set breath rate, whichever is greater; Low O2 (VSO2 only): 35 PSIG (2.4 Bar); High O2 (VSO2 only): 65 PSIG (4.3 Bar); Ventilator Inoperative: Red Indicator; Alarm Silence: 60 second max.; Alarm Volume: 50 - 70 dba | ||
| Non-Adjustable Alarms: | Alarms: Low Pressure/Apnea; High Pressure; Setting Error; Power Switchover; Low Power; O2 Failure Alerts: External Battery; Internal Battery; Battery Charge; Alarm Control; Assist/Spontaneous | ||
Monitored Variables : | |||
| Pressure | Airway Pressure Manometer: -20 to 120 cmH2O; Peak Inspiratory Pressure (PIP): 0 to 140 cmH2O; Mean Airway Pressure (MAP): 0 to 99 cmH2O; Positive End Expiratory Pressure (PEEP); cmH2O: 0 to 99 cmH2O | ||
| Respiratory Rate | Total Breath Rate (f): 0 to 250 bpm | ||
| Tidal Volume | 0 to 4000 ml | ||
| Minute Volume | 0 to 99.9 L | ||