| Puritan-Bennett Infant Star 100 Ventilator System | |||
Basic Description | |||
| Manufacturer: | Puritan Bennett Corporation | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) Transport CPAP | ||
| Patient Population Range: | Neonatal Pediatric | ||
Characteristics | |||
| Operating Range of Controls: | Inspiratory Time: 0.2 to 30 seconds; Expiratory Time: 0.2 to 30 seconds; Ventilator Rate: 2 to 150 bpm determined from inspiratory and expiratory time settings; Flow: 0 to 15 Lpm; PEEP/CPAP: 0 to 60 cmH2O; Peak Inspiratory Pressure: 0 to 60 cmH20 | ||
| Size: | 7.25" H x 8.5" W x 4" D (18 x 22 x 4 cm) | ||
| Weight: | 7 lb (3.2 kg) | ||
| Other Characteristics: | Tracking Relief Valve: Limits airway pressure at 10 cmH20 above PIP; Internal Safety Relief: Limits breathing circuit from excessive subambiant pressure (-2 to -5 cmH2o) and excessive positive pressure (72 to 92 cmH20) | ||
Characteristics (Non - Technical) | |||
| Warranty Period: | One year (parts and labor) | ||
| Other Comments: | Breathing circuit, exhalation valve (reusable), high pressure hose(s), and 115 V or 230 V battery charger/power supply are included with each ventilator. | ||
Specifications | |||
| Front Panel Controls: | nspiratory Time; Expiratory Time; Ventilator Rate; Flow; PEEP/CPAP; PIP (Peak Inspiratory Pressure); Manual Breath; Proximal Airway Pressure | ||
| Power Requirements: | Electrical Power Internal Battery Operation: Rechargable gel-cell battery offering 6 to 8 hours of continuous operation; Battery Charger/Power Supply (included): 120 VAC, 60 Hz, 0.18 amp; 230 VAC, 50/60 Hz, 0.125 amp Allows recharging during ventilator operations/standby Pneumatic Power Internal: 40 to 70 psig (276 to 483 kPa) blended Air/O2; Bleed flow: 6 Lpm (maximum) | ||
| Languages: | English, Russian, Japanese | ||
| Alarms: | General System Failure; Additional General System Alarms; Low Peak Pressure; Low PEEP; Apnea (optional) | ||
| Alarm Controls: | Low Pressure Alarm Delay: 0 to 40 seconds; Proximal Airway Pressure: -20 to 100 cmH20 | ||
| Non-Adjustable Alarms: | Low Pressure Alarm Delay: 0 to 40 seconds; Proximal Airway Pressure: -20 to 100 cmH20 | ||
| AC Power and Battery Indicators: | Amber with external AC power | ||
| Other Specifications: | Other indicators: Battery Low Alarm: flashes red when charging is required; Inspiratory LED: glows amber during mandatory inspiratory phase; Battery Charging: indicates battery charging status | ||