Hamilton GALILEO Ventilator
   

Basic Description

 
 
Manufacturer:  Hamilton Medical AG
Ventilator Type:  Positive Pressure
Invasive
Conventional
Nitric Oxide
Ventilator Class:  Intensive Care Unit (ICU)
Transport
CPAP
Patient Population Range:  Neonatal
Pediatric
Adult
 

Description

 
Features:  Adaptive Support Ventilation (ASV) maintains an operator-preset, minimal Minute Ventilation independent of the patient's activity.
 

Characteristics

 
Modes--Invasive:  Volume-controlled modes --- ASSIST/CONTROL; SIMV (PSV); MMV; SPONT; (S)CMV; ASV
Pressure-controlled modes --- ASSIST/CONTROL; SIMV (PSV); CPAP; APV
Waveforms:  Sine, square, decelerating, 50% decelerating
Operating Range of Controls:  Special functions: Manual breath, inspiratory/expiratory hold, nebulizer, 100% O2, sigh, apnea backup; CMV rate: 5 to 120 b/min; SIMV rate: 1 to 60 b/min; Tidal volume: 10 to 2000 ml; PEEP/CPAP 0 to 50 cmH2O; Oxygen: 21 to 100%; I:E ratio: 1:9 to 4:1; Inspiratory time: 10 to 80% of cycle time (0.1 to 3 s); Pause time 0 to 70% of cycle time (0 to 8 s); Peak flow 1 to 180 l/min; Pressure trigger 0.5 to 10 cmH2O below PEEP/CPAP; Flow trigger 0.5 to 15 l/min; Base flow 4 to 30 l/min, depending on flow trigger setting; Pressure control 5 to 100 cmH2O; Pressure support 0 to 100 cmH2O; Pressure ramp 25 to 200 ms TO..63; Expiratory trigger sensitivity 10 to 40% of inspiratory peak flow; % minute volume (ASV) 10 to 350%; Bodyweight (ASV) 10 to 200 kg
Triggering:  Pressure trigger: 0.5 to 10 cm H2O below PEEP/CPAP. Flow trigger: 0.5 to 15 l/min.
Size:  17.5 W x 24.5 D x 61 H
Weight:  97 lbs.
Other Characteristics:  Monitoring: Real-time waveforms -- volume, flow, airway pressure, auxiliary pressure; Pressures -- peak, mean, minimum, plateau, PEEP/CPAP, autoPEEP, P(01); Flow -- inspiratory peak, expiratory peak; Volume -- tidal, minute, leakage; Time -- inspiratory, expiratory, I:E ratio, total frequency, spontaneous frequency; Others -- Inspiratory resistance, expiratory resistance, static compliance, inspiratory time constant, expiratory time constant, imposed work of breathing, pressure time product, rapid shallow breathing index, delivered oxygen concentration; Options: Nebulizer, communications interface including RS-232C port, remote nurse’s call, I:E signal

Characteristics (Non - Technical)

Approvals:  IEC 601-1/EN 60601-1, IEC 601-1-2/EN 60601-1-2, EN 794-1, ISO 7767. CE. CSA, FDA

Specifications

Power Requirements:  Electrical requirements: 100 to 240 V ~ +/-10%, 50/60 Hz; O2 and air supply: 200 to 600 kP a (29 to 86 psi)
Alarms:  Low/high minute volume: 0 to 50 l/min; Low/high pressure: 0 to 110 cmH2O; Low/high tidal volume: 0 to 3000 ml; Low/high rate: 0 to 130 b/min; Apnea time: 10 to 60 s; Air trapping: 1 to 10 l/min; Special: Oxygen concentration, disconnection, oxygen/air supply, loss of PEEP, check settings, flow sensor alarms, power supply, inoperative
Mounting Options:  Standard trolley or shelf mount
Other Specifications:  Operating environment: Temperature – 10 to 40 C, (-10 to 60 C during transport); Humidity -- 30 to 75% noncondensing, (5 to 85% noncondensing during transport); Altitude – 3000 m; Backup battery time: 20 minutes minimum (fully charged)

Monitored Variables :

Tidal Volume 10 to 2000 ml

Related Info 

Distributor |Educator | Manufacturer |Rental Agent | Service Center |
Parts Available | Product Markets | Service Centres |