| Respironics PLV-100 Ventilator | |||
Basic Description | |||
| Manufacturer: | Respironics | ||
| Ventilator Type: | Positive Pressure | ||
| Ventilator Class: | Transport Home | ||
| Patient Population Range: | Pediatric Adult Long-term (chronic) | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- CONTROL; ASSIST/CONTROL; SIMV Pressure-controlled modes --- CONTROL; ASSIST/CONTROL; SIMV | ||
| Waveforms: | Sine Wave Flow With Sigmoidal Pressure Pattern | ||
| Operating Range of Controls: | Tidal Volume: 50 to 3000 ml; Rate: 2 to 40 per min.; Peak Flow Rate: 10 to 120 liters/min.; Sensitivity: -6 to +3 cmH2O | ||
| Size: | 22.9 x 31.1 x 31.1 (cm) | ||
| Weight: | 12.8 (kg) | ||
Characteristics (Non - Technical) | |||
| List Price: | $12350 | ||
| Warranty Period: | 1 year | ||
| Other Comments: | Monthly maintenance contract. Was formerly manufactured by Lifecare. | ||
Specifications | |||
| Front Panel Controls: | Mode Selection: Control, Assist/Control, SIMV; Tidal Volume: 0.05 to 3.00 liters +/- 10%; Breath Rate: 2 to 35 BPM +/- 0.5; 36 to 40 BPM +/- 2; Inspiratory Flow Rate*: 10 to 120 L/Min; Battery Test Switch: Test Internal/External Battery; Main Power Switch: On/Off, Recharge; 5 Amp Circuit Breaker: Push/Pull Actuation; 2 Amp Circuit Breaker: Push/Pull Actuation; Airway Pressure Limit: 5 to 95 cm H2O +/- 5; Patient Airway Pressure**: -10 to 100 cm H2O +/- 5; Sensitivity: +3 cm H2O +/- 1 to at least +/- 6 cm H2O; Low Pressure Alarm: 2 to 40 cm H2O +/- 2; 41 to 50 cm H2O +/- 5; (* Peak flow rate during inspiratory time); (** Pressure waveform is sinusoidal). | ||
| Display Monitor Controls: | Tidal Volume: Digital Display; Rate: Patient Breaths Per Minute: Digital Display; I:E Ratio: Digital Display; Inspiratory Flow Rate: Digital Display; Increase Inspiratory Flow: Red Indicator Light; AC Power Source: Green Indicator Light; Internal DC Power Source: Yellow Indicator Light; External DC Power Source: Amber Indicator Light; Patient Airway Pressure: Gauge; Assisted/Spontaneous Breath: Green Indicator Light; 15-Second Delay (Low Pressure Alarm): Green Indicator Light. | ||
| Alarms: | Low Pressure: Visual/Continuous Audible; High Pressure: Intermittent Audible; Apnea: Visual/Continuous Audible; Inverse I:E Ratio: Visual Warning; Increase Inspiratory Flow: Visual Warning; Low Internal Battery: Visual/Continuous Audible; Low External Battery: Visual/Continuous Audible; Reverse Battery Connection: Continuous Audible; Power Failure: Continuous Audible; Power Source: Audible; Unit Malfunction: Intermittent Audible | ||
| Other Specifications: | Internal battery duration 1 hour, external battery duration 24 hours | ||