| Puritan-Bennett LP10 Ventilator | |||
Basic Description | |||
| Manufacturer: | Puritan Bennett Corporation | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) Transport Subacute Home | ||
| Patient Population Range: | Pediatric Adult Acute Long-term (chronic) | ||
Description | |||
| Brief Description: | Go Beyond The Static Specs! Click Here For Special Online Interactive Content! (Requires Flash Player 5.0) | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- CONTROL; ASSIST/CONTROL; SIMV Pressure-controlled modes --- CONTROL; ASSIST/CONTROL; SIMV | ||
| Waveforms: | Sigmoidal Pressure Pattern | ||
| Operating Range of Controls: | Pressure Limit: Locking knob sets pressure from 15 to 50 cmH2O/hPa, or disables limit feature; Volume: Push-to-turn knob: continuously adjustable from 100 to 2200 ml; Breath Rate: Set breathing rate: 1 to 20 BPM in increments of 1 BPM and 22 to 38 BPM in increments of 2 BPM; Inspiratory Time: Set time from 0.5 to 1.0 sec in increments of 0.1 sec, 1.2 sec, and 1.5 to 5.5 sec in increments of 0.5 sec; Breathing Effort: Push-to-turn knob sets pressure level to detect breathing effort (-10cmH2O/hPa to +10cmH2O/hPa) | ||
| Size: | 24.6 x 36.8 x 33.6 (cm) | ||
| Weight: | 15.9 (kg) | ||
| Power: | AC 110V/60Hz | ||
| Other Characteristics: | Operating Environment: 5°C to 40°C (41°F to 104°F), 10% to 90% RH non-condensing Storage Environment: -15°C to 50°C (5°F to 122°F), 10% to 90% RH non-condensing; allow at least 1 hour temperature stabilization before use when moving from non-operating to operating environment Indicators, Normal Events Patient Pressure Meter: Displays patient pressure -10 to +100 cmH2O/hPa; can also display battery charge and machine hours of operation; Breathing Effort: Green LED activated by adequate patient breathing effort; Power: LEDs indicate operating power source green: AC Pwr/Batt Chg; amber: External Battery; flashing amber: Internal Battery; Resistance Factor: Less than 0.05 cmH2O/hPa/LPM; Emergency Pressure Relief: Approx. 100 cmH2O/hPa; Flow: 20-100 LPM | ||
Characteristics (Non - Technical) | |||
| List Price: | $9950 | ||
| Other Comments: | Originally manufactured by Aequitron Medical, now owned by Mallinckrodt. Recommend 6000 hr. preventative maintenance or 12 month recertification. | ||
Specifications | |||
| Front Panel Controls: | Front Panel Touch Pads Alarm Silence/Reset: Push button to silence alarm or to reset; used with Battery Test button to read operating hours on Patient Pressure meter; Battery Test: Push button to show charge level on lower window of Patient Pressure meter; Front Panel Controls Mode: Standby, Assist/Control, SIMV or Pressure Cycle; Pressure Limit; Low Alarm; High Alarm Limit; Volume; Breath Rate; Inspiratory Time; Breathing Effort | ||
| Power Requirements: | 110 VAC range (100-127 VAC) or 220 VAC range (220-240 VAC), 50/60 Hz, external voltage selector switch | ||
| Alarm Controls: | Low Alarm: Set limit for Low Pressure alarm (2-32 cmH2O/hPa in increments of 2 cmH2O/hPa); High Alarm Limit: Set limit for High Pressure alarm (Assist/Control, SIMV Modes) or limit for Pressure Cycle mode (15-90 cm H2O/hPa in increments of 5 cmH2O/hPa) | ||
| Non-Adjustable Alarms: | Low Pressure/Apnea: Pulsating tone, flashing red LED; Low Power: Pulsating tone, flashing red LED; High Pressure: Pulsating tone, flashing red LED; Setting Error: Pulsating tone, flashing red LED; Power Switchover: Pulsating tone, flashing red LED; Loss of Microprocessor Control: Steady tone; Every 5 Minutes When Powered by Internal Battery: Reminder tone; Each Time Accessory Printer Generates Report: Reminder tone | ||
| Interface Ports: | Rear connection for optional printer and digital display | ||
| Other Specifications: | Power Usage Maximum: 630 kW hours per year Nominal: 315 kW hours per year External Battery: Approximately 20 hours operation with 75-80 ampere-hour 12 VDC deep-cycle, gel-cell battery; approximately 10 hours with 35-40 ampere-hour 12 VDC deep-cycle, gel-cell battery Internal Battery: Up to 1 hour operation Alarms classified by urgency. | ||
Monitored Variables : | |||
| Pressure | -10 to +100 cmH2O/hPa | ||
Related Info | |||