| Dräger Oxylog 2000 Ventilator | |||
Basic Description | |||
| Manufacturer: | Dräger Medical, Inc. | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Transport | ||
| Patient Population Range: | Adult | ||
Characteristics | |||
| Modes--Invasive: | Time-cycled modes --- CMV, SIMV/CPAP | ||
| Waveforms: | Flow: Rectangular | ||
| Size: | 21 x 12 x 20.5 (cm); Battery 4 1.5 AA(?) | ||
| Weight: | 4.3 (kg); Battery 0.1 (kg) | ||
Specifications | |||
| Alarms: | Disconnection; Stenosis; Apnea; Paw High; Incorrect Settings | ||
Monitored Variables : | |||
| Pressure | -10 to 80 cmH2OPMAX: 20 to 60 cmH2O; Ventilator rate: 5 to 40 cycles/minute; Flow limit: 4 to 60 L/min; Tidal volume: 0.1 to 1.5 L; Inspiratory time: CMV: 0.5 to 2.0 seconds; SIMV: 5 to 12 bpm: 2.0 seconds; 12 to 40 bpm: whatever I:E of 1:1.5 permits (0.5 to 2.0 seconds); Baseline pressure: 0 to 15 cmH2O (aneroid gauge) | ||
| Minute Volume | 2 to 40 L/min (calculated from flow measurement) | ||
Related Info | |||