Siemens Servo 300 Ventilator
   

Basic Description

 
 
Manufacturer:  MAQUET GmbH & Co. KG
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Transport
CPAP
Patient Population Range:  Neonatal
Pediatric
Adult
 

Description

 
Features:  Customizable respiratory patterns, supported and controlled ventilation modes, easy setup using context-sensitive controls, automatic emergency battery backup.
Brief Description:  Go Beyond the Static Specs! Click Here For Special Online Interactive Content! (Requires Flash Player 5.0)
 

Characteristics

 
Modes--Invasive:  Pressure-controlled modes --- PRESSURE CONTROL, PRESSURE SUPPORT, SIMV (PRESSURE CONTROL) + PRESSURE SUPPORT, PRESSURE REGULATED VOLUME CONTROL, VOLUME SUPPORT, CPAP
Flow-controlled modes --- VOLUME CONTROL; SIMV (VOLUME CONTROL) + PRESSSURE SUPPPORT
Waveforms:  Pressure: Rectangular, Adjustable Slope; Flow: Rectangular, Adjustable Slope
Operating Range of Controls:  Trigger Sensitivity Level: Pressure: (0 to 20 cmH2O below PEEP); Flow: adult: 0.6 - 2.0 L/min, pediatric: 0.3 - 1.0 L/min, neonate: 0.15 - 0.5 L/min; CMV Frequency: 5 to 150 breaths/min; SIMV Frequency: 0.5 to 40 breaths/min; Pressure Control Level Above PEEP: 0 to 100 cmH2O; Pressure Support Level Above PEEP: 0 to 100 cmH2O; Pause Time: 0% to 30%; Minute Volume: 0.2 to 60 L/min; Inspiratory Time: 10% to 80% of the ventilatory period; Baseline Pressure: 0 to 50 cmH2O
Triggering:  Pressure, flow, time, manual
Graphic Display:  Red Digital Display;
Power:  AC 110V/60Hz
Other Characteristics:  Also monitors I:E ratio

Specifications

Front Panel Controls:  Patient type (adult, pediatric, neonate); Mode of ventilation; Pressure or flow triggering; Pressure limit; Baseline pressure; Minute volume; Ventilator frequency; Start breath; Pause hold; Inspiratory time (%); Inspiratory pause time (%); Rise time of inspiratory pressure; FIO2; Deliver preset group of 100% oxygen breaths
Power Requirements:  Electrical power: 115 VAC, 60 Hz; Pneumatic power: Air and oxygen at 29 to 100 psig
Alarm Controls:  Airway pressure: Pressure > upper pressure limit or set PEEP level plus 15 cmH2O for more than 15 seconds; O2 concentration: FlO2 above or below 6% of the set alarm value or if the oxygen fuel cell is not connected; Exp minute volume alarm: minute volume > set value or < 0.3 L/min for adults and children or 0.06 L/min for neonates.
Non-Adjustable Alarms:  Battery: main electric power has failed and the ventilator has switched to battery backup; Gas Supply: gas pressure < 29 psig or > than 94 psig; Technical: a wide variety of technical problems, most of which require technical service; Apnea: time between breaths > 10 seconds for adults, 15 seconds for children, or 20 seconds for neonates
Other Specifications:  30 minute battery duration. Alarms classified by urgency.

Monitored Variables :

Pressure peak, mean, pause, baseline, and an analog bar graph of instantaneous airway pressure
Respiratory Rate
Tidal Volume inspired and expired
Minute Volume

Related Info 

Distributor |Educator | Manufacturer |Rental Agent |Service Center |
Parts Available | Product Markets | Service Centres |

Distributor:

Freedom Medical
208 Carter Drive
West Chester, PA



Manufacturer:

MAQUET GmbH & Co. KG
Kehlerstraße 31
Rastatt,
Germany



Rental Agent:

Freedom Medical
208 Carter Drive
West Chester, PA