Siemens Servo 900C Ventilator
   

Basic Description

 
 
Manufacturer:  MAQUET GmbH & Co. KG
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Transport
Subacute
Patient Population Range:  Neonatal
Pediatric
Adult
 

Description

 
Features:  It can be used during MRI scans, synchronization of multiple units allows for ventilation of both lungs individually.
 

Characteristics

 
Modes--Invasive:  Pressure-controlled modes --- SIMV (spontaneous breaths); SIMV + PRESSURE SUPPORT; PRESSURE CONTROL; PRESSURE SUPPORT; CPAP
Flow-controlled modes --- VOLUME CONTROL; VOLUME CONTROL SIGH; SIMV (mandatory breaths)
Waveforms:  Pressure: Rectangular; Flow: Rectangular, Ascending-Descending ramp
Operating Range of Controls:  Trigger Sensitivity Below PEEP: 0 to 20 cmH2O; Frequency: 10 - 120 breaths/min; SIMV breaths: Upper range: 4 - 40 breaths/min; Lower range: 0.5 - 4 breaths/min; Upper Pressure Limit: 0 to 100 cmH2O above set PEEP; PEEP: - 10 to 50 cmH2O (NEEP requires a vacuum source connected to the expiratory side of the patient circuit; Inspiratory Pressure Limit Above PEEP: 16 to 120 cmH2O above ambient; Inspiratory Flow Limit: 0.5 to 40 L/min; Inspiratory Time %: 20%, 25%, 33%, 50%, 67%, or 80%; Pause Time % (Inspiratory Pause): 0% to 30%
Triggering:  Pressure, time
Power:  AC 110V/60Hz
Other Characteristics:  Also displays inspired oxygen concentration. Two Servo 900C ventilators can be synchronized to ventilate each of a patient's lungs differently.

Specifications

Front Panel Controls:  Mode; Minute Volume; Trigger Sensitivity Below PEEP; Upper Pressure Limit; PEEP; Inspiratory Pressure Limit Above PEEP; Frequency (with range toggle); Inspiratory Flow Limit and Waveform; Inspiratory Time %; Pause Time % (Inspiratory Pause)
Alarm Controls:  High pressure: 20 to 100 cmH2O; Low pressure: 20 to 100 cmH2O; High expired minute volume: 3 to 43 L/min (adults) or 0 to 4.3 L/min (infants); Low expired minute volume: 0 to 37 L/min (adults) or 0 to 3.7 L/min (infants); High O2 %: 30% to 100%; Low O2 %: 18% to 90%
Non-Adjustable Alarms:  Interrupted power supply; Inadequate gas supply; Apnea: no attempt to inspire for 15 seconds

Monitored Variables :

Pressure PIP, Pause pressure, mean airway pressure, ventilating pressure, working pressure
Tidal Volume inspired and expired
Minute Volume expired

Related Info 

Distributor |Educator | Manufacturer |Rental Agent |Service Center |
Parts Available | Product Markets | Service Centres |

Distributor:

Freedom Medical
208 Carter Drive
West Chester, PA



Distributor:

Pioneer Medical, Inc.
566 Mainstream Drive
Nashville, TN



Manufacturer:

MAQUET GmbH & Co. KG
Kehlerstraße 31
Rastatt,
Germany



Rental Agent:

Freedom Medical
208 Carter Drive
West Chester, PA