| Sechrist IV-200 SAVI Ventilator | |||
Basic Description | |||
| Manufacturer: | Sechrist Industries | ||
| Ventilator Type: | Positive Pressure Invasive Conventional High Frequency | ||
| Ventilator Class: | Intensive Care Unit (ICU) | ||
| Patient Population Range: | Neonatal Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Pressure-controlled modes --- IMV (available with SAVI) Supplemental modes --- CPAP (not controlled) General modes --- CMV | ||
| Waveforms: | Pressure: Rectangular; Flow: Rectangular | ||
| Operating Range of Controls: | Flow: 0 to 32 LPM; Oxygen Percent: 21 to 100%; Inspiratory Time: 0.10 to 2.90 sec; Expiratory Time: 0.30 to 60.0 sec; Inspiratory Pressure: 5 to 70 cm H20; Expiratory Pressure: 2 to 20 cm H20 | ||
| Size: | 41.27 x 30.48 x 24.13 (cm) | ||
| Weight: | 12.2 (kg) | ||
| Power: | |||
| Other Characteristics: | Safety Pressure Relief Valve (Adjustable): 15 to 85 cm H20 | ||
Specifications | |||
| Power Requirements: | 100-120 VAC 50/60 Hz; 220-240 VAC 50/60 Hz; Circuit Breaker Protected; D.C. Power Requirements:12 VDC 1.2 amps; | ||
| Alarms: | Alarms High (HI) Pressure Limit Peak Overpressure; Low (LO) Pressure Limit; Low Airway Pressure; Leaks, Patient Disconnect; Fail-to-Cycle; Apnea; Prolonged Inspiration; Source Gas Failure; Power Failure (Battery Backup) | ||
| Alarm Controls: | Delay Time: 3 to 60 sec.; Alarm Mute: 30 +/- 5 sec. | ||
Monitored Variables : | |||
| Pressure | -5 to 80 cm H20 | ||
| Respiratory Rate | |||