| Puritan-Bennett 7200ae Ventilator | |||
Basic Description | |||
| Manufacturer: | Puritan Bennett Corporation | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) Subacute CPAP | ||
| Patient Population Range: | Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Pressure-controlled modes --- PRESSURE CONTROL; SIMV (spontaneous breaths); CPAP (spontaneous inspirations); (requires Option 80 software) Flow-controlled modes --- CMV; FLOW-BY Time-cycled modes --- SIMV; CMV Combined modes --- SIMV (mandatory breaths) (pressure/flow controlled) Supplemental modes --- Disconnect ventilation, Back-up ventilation (for details, see Other Characteristics) | ||
| Waveforms: | Pressure: Rectangular; Flow: Rectangular, Descending Ramp, Sinusoidal | ||
| Operating Range of Controls: | Respiratory Rate: 0.5 to 70 bpm; Tidal Volume: 0.10 to 2.5 liters; Peak Flow: 10 to 120 Lpm; O2%: 21% to 100%; Sensitivity: 0.5 to 20.0 cmH2O. If Flow-by is active then flow sensitivity ranges from 1 to 15 Lpm; Plateau: 0.0 to 2.0 seconds; Inspiratory Pressure: 5 to 100 cmH2O above PEEP; I:E ratio: Inverse ratio up to 4:1. Conventional ratio up to 1:9; Inspiratory Time: 0.2 to 9.9 seconds; PEEP/CPAP: 0 to 45 cmH2O; Pressure Support: 1 to 70 cm H2O; Sigh Volume: 0 to 2.5 liters (no more than two times the Tidal Volume); Sigh Pressure Limit: 10 to 120 cmH2O; Sigh Rate: 1 to 15 events per hour; Multiple Sighs: Adjustable in multiples of 1, 2, or 3; (Some of the settings are mode-specific.) | ||
| Graphic Display: | Tidal Volume; Set Rate (bpm); Peak Flow Lpm: O2%; PEEP/CPAP | ||
| Size: | 16" x 22" x23" | ||
| Weight: | 145 pounds | ||
| Power: | AC 110V/60Hz | ||
| Other Characteristics: | Has a back-up preprogrammed analog ventilator for some emergency situations. (Tidal Volume: 0.5 liters; Rate: 12 bpm; Peak flow: 45 Lpm; Sensitivity: lockout; O2%: 100% or 21%, depending on condition of high pressure supplies; High Pressure Limit: PEEP proportional pressure + 30 cmH2O. If all gas supplies are lost, during a Power-On Self Test, a system fault is detected, or AC power is lost, the ventilator goes into Safety Valve Open (SVO) mode. The expiratory and inspiratory limbs of the patient circuit are opened to room air. Rebreathing of exhaled gases are prevented by check valves. | ||
Characteristics (Non - Technical) | |||
| Warranty Period: | 2 years | ||
| Other Comments: | 5 year contracts available, 10,000 hour major preventative maintenance interval | ||
Specifications | |||
| Front Panel Controls: | Apnea parameters; Auto PEEP; Pressure Support; DCI 2.0; Respiratory Mechanics; FlowBy 2.0; Graphics 2.0; Pressure Control; Tidal Volume; Respiratory Rate; Peak Inspiratory Flow; Sensitivity; O2%; Plateau; High Pressure Limit; Low Inspiration Pressure; Low PEEP/CPAP Pressure; Low Exhaled Tidal Volume; Low Exhaled Minute Volume; High Respiratory Rate; CMV; SIMV; CPAP; Square Wave; Descending RampWave; Sine Wave; Manual Inspiration; Manual Sigh; Automatic Sigh; 100% O2 Suction; Nebulizer | ||
| Power Requirements: | Dry wall air or O2 at pressure from 35 psig to 100 psig. | ||
| Alarms: | General System Failure; Inverse I:E Ratio; Additional General System Alarms; Low Peak Pressure; Low PEEP; Low Tidal Volume; Low Minute Ventilation; High/Low Breath Rate; Apnea | ||
| Alarm Controls: | High Pressure Limit: 10 to 120 cmH2O; Low Inspiration Pressure: 3 to 99 cmH2O; Low PEEP/CPAP Pressure: 0 to 45 cmH2O; Low Exhaled Tidal Volume: 0.0 to 2.59 liters; Low Exhaled Minute Volume: 0.0 to 60.0 liters; High Respiratory Rate: 0 to 70 bpm | ||
Monitored Variables : | |||
| Pressure | Vertical LED bar graph of tubing airway pressure: -10 to 60 cmH2O, -20 to 120 cmH2O, digital display of mean airway pressure, peak airway pressure, or plateau pressure (one at a time) | ||
| Respiratory Rate | Displays breaths per minute or I:E ratio | ||
| Minute Volume | Tidal volume, minute volume, or spontaneous minute volume (one at a time). These are corrected for tubing compliance and body temperature pressure saturated (BTPS). | ||