Puritan-Bennett 7200ae Ventilator
   

Basic Description

 
 
Manufacturer:  Puritan Bennett Corporation
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Subacute
CPAP
Patient Population Range:  Pediatric
Adult
 

Characteristics

 
Modes--Invasive:  Pressure-controlled modes --- PRESSURE CONTROL; SIMV (spontaneous breaths); CPAP (spontaneous inspirations); (requires Option 80 software)
Flow-controlled modes --- CMV; FLOW-BY
Time-cycled modes --- SIMV; CMV
Combined modes --- SIMV (mandatory breaths) (pressure/flow controlled)
Supplemental modes --- Disconnect ventilation, Back-up ventilation (for details, see Other Characteristics)
Waveforms:  Pressure: Rectangular; Flow: Rectangular, Descending Ramp, Sinusoidal
Operating Range of Controls:  Respiratory Rate: 0.5 to 70 bpm; Tidal Volume: 0.10 to 2.5 liters; Peak Flow: 10 to 120 Lpm; O2%: 21% to 100%; Sensitivity: 0.5 to 20.0 cmH2O. If Flow-by is active then flow sensitivity ranges from 1 to 15 Lpm; Plateau: 0.0 to 2.0 seconds; Inspiratory Pressure: 5 to 100 cmH2O above PEEP; I:E ratio: Inverse ratio up to 4:1. Conventional ratio up to 1:9; Inspiratory Time: 0.2 to 9.9 seconds; PEEP/CPAP: 0 to 45 cmH2O; Pressure Support: 1 to 70 cm H2O; Sigh Volume: 0 to 2.5 liters (no more than two times the Tidal Volume); Sigh Pressure Limit: 10 to 120 cmH2O; Sigh Rate: 1 to 15 events per hour; Multiple Sighs: Adjustable in multiples of 1, 2, or 3; (Some of the settings are mode-specific.)
Graphic Display:  Tidal Volume; Set Rate (bpm); Peak Flow Lpm: O2%; PEEP/CPAP
Size:  16" x 22" x23"
Weight:  145 pounds
Power:  AC 110V/60Hz
Other Characteristics:  Has a back-up preprogrammed analog ventilator for some emergency situations. (Tidal Volume: 0.5 liters; Rate: 12 bpm; Peak flow: 45 Lpm; Sensitivity: lockout; O2%: 100% or 21%, depending on condition of high pressure supplies; High Pressure Limit: PEEP proportional pressure + 30 cmH2O. If all gas supplies are lost, during a Power-On Self Test, a system fault is detected, or AC power is lost, the ventilator goes into Safety Valve Open (SVO) mode. The expiratory and inspiratory limbs of the patient circuit are opened to room air. Rebreathing of exhaled gases are prevented by check valves.

Characteristics (Non - Technical)

Warranty Period:  2 years
Other Comments:  5 year contracts available, 10,000 hour major preventative maintenance interval

Specifications

Front Panel Controls:  Apnea parameters; Auto PEEP; Pressure Support; DCI 2.0; Respiratory Mechanics; FlowBy 2.0; Graphics 2.0; Pressure Control; Tidal Volume; Respiratory Rate; Peak Inspiratory Flow; Sensitivity; O2%; Plateau; High Pressure Limit; Low Inspiration Pressure; Low PEEP/CPAP Pressure; Low Exhaled Tidal Volume; Low Exhaled Minute Volume; High Respiratory Rate; CMV; SIMV; CPAP; Square Wave; Descending RampWave; Sine Wave; Manual Inspiration; Manual Sigh; Automatic Sigh; 100% O2 Suction; Nebulizer
Power Requirements:  Dry wall air or O2 at pressure from 35 psig to 100 psig.
Alarms:  General System Failure; Inverse I:E Ratio; Additional General System Alarms; Low Peak Pressure; Low PEEP; Low Tidal Volume; Low Minute Ventilation; High/Low Breath Rate; Apnea
Alarm Controls:  High Pressure Limit: 10 to 120 cmH2O; Low Inspiration Pressure: 3 to 99 cmH2O; Low PEEP/CPAP Pressure: 0 to 45 cmH2O; Low Exhaled Tidal Volume: 0.0 to 2.59 liters; Low Exhaled Minute Volume: 0.0 to 60.0 liters; High Respiratory Rate: 0 to 70 bpm

Monitored Variables :

Pressure Vertical LED bar graph of tubing airway pressure: -10 to 60 cmH2O, -20 to 120 cmH2O, digital display of mean airway pressure, peak airway pressure, or plateau pressure (one at a time)
Respiratory Rate Displays breaths per minute or I:E ratio
Minute Volume Tidal volume, minute volume, or spontaneous minute volume (one at a time). These are corrected for tubing compliance and body temperature pressure saturated (BTPS).

Related Info 

Distributor |Educator | Manufacturer |Rental Agent |Service Center |
Parts Available | Product Markets | Service Centres |

Distributor:

Huff and Puff Victoria Inc.
PO Box 168
Melbourne,