Puritan-Bennett Infant Star 200 Ventilator
   

Basic Description

 
 
Manufacturer:  Puritan Bennett Corporation
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Subacute
CPAP
Patient Population Range:  Neonatal
Pediatric
Acute
Long-term (chronic)
 

Characteristics

 
Modes--Invasive:  Pressure-controlled modes --- ASSIST/CONTROL; SIMV; CPAP (these require Star Sync); IMV; CMV
Supplemental modes --- CPAP (not controlled)
Waveforms:  Pressure: Rectangular; Flow: Rectangular
Operating Range of Controls:  Ventilator Rate: 1 to 150 bpm; Inspiratory Time: 0.1 to 2.9 seconds; Flow: 0 to 50 Lpm; PEEP/CPAP: 0 to 20 cmH20; PIP (Peak Inspiratory Pressure): 0 to 70 cmH2O; I:E Ratio: 1:0.25 to 1:9.9; Mean Airway Pressure: 0 to 99.9 cmH2O
Size:  9" (23cm) H x 15.5" (39cm) W x 10" (25cm) D
Weight:  17 lb (7.7 kg) without blender; 20 lb (20 kg) with blender
Other Characteristics:  Safety features: Tracking relief valve: Limits airway pressure at 10 cm H2O above PIP; Internal safety relief: Limits breathing circuit from excessive subambiant pressure (-2 to -5 cmH2O) and excessive positive pressure (72 to 92 cmH2O)

Characteristics (Non - Technical)

Other Comments:  Each ventilator comes with breathing circuit, reusable exhalation valve, high pressure hoses, and test lung.

Specifications

Front Panel Controls:  Ventilator Rate; Inspiratory Time; Flow; PEEP/CPAP; PIP (Peak Inspiratory Pressure); Manual Breath (delivers single mandatory breath); I:E Ratio; Mean Airway Pressure
Power Requirements:  Electrical Power: 108 to 132 VAC, 60 Hz; 207 to 242 VAC, 50/60 Hz or 12 VDC, 2.5 amp; Pneumatic Power: Internal: 40 to 70 psig (275 to 481 kPa); Bleed Flow: 4 to 7 Lpm w/o blender, 12 to 15 Lpm w/blender
Alarms:  Low Pressure Alarm Silence: 60 seconds
Alarm Controls:  Proximal Airway Pressure: -20 to 100 cmH2O; High Pressure Alarm: 1 to 85 cmH2O; Low Pressure Alarm: 1 to 85 cmH2O
Non-Adjustable Alarms:  Insufficient Expiratory Time: Preset at 0.2 seconds; Low Inlet Pressure: Preset for 40 psig inlet pressure; Max Inverse I:E Ratio: Preset for ratios >4:1; Power Loss: Insufficient AC or DC power; Ventilator Inop: Malfunction detected by microprocessor
Mounting Options:  Star cart (with or without electrical outlet), pedestal

Monitored Variables :

Pressure Mean airway pressure (digital), airway pressure (aneroid gauge)