Nellcor Puritan-Bennett Infant Star Ventilator
(Discontinued)
   

Basic Description

 
 
Manufacturer:  Puritan Bennett Corporation
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Patient Population Range:  Neonatal
 

Characteristics

 
Modes--Invasive:  Pressure-controlled modes --- CONTINUOUS-FLOW IMV; DEMAND-FLOW IMV; CONTINUOUS-FLOW CPAP; DEMAND-FLOW CPAP
Waveforms:  Pressure-limited: Rectangular; Flow: Rectangular.
Operating Range of Controls:  Frequency: 1 to 150 breaths/minute (1 breath increments from 1 to 60, 2 breath increments from 61 to 130, 5 breath increments from 131 to 150); Pressure limit: 5 to 90 cmH2O; Continuous flow: 4 to 40 L/min; Inspiratory time: 0.1 to 3.0 seconds; Baseline pressure (PEEP/CPAP): 0 to 24 cmH2O
Triggering:  Time
Other Characteristics:  As a safety feature, knobs must be depressed before they can be turned.

Specifications

Front Panel Controls:  Inspiratory Time; Expiratory Time; Ventilator Rate; Flow; PEEP/CPAP; PIP (Peak Inspiratory Pressure); Manual Breath; Proximal Airway Pressure
Power Requirements:  Electrical Power Internal Battery Operation: Rechargable gel-cell battery offering 6 to 8 hours of continuous operation; Battery Charger/Power Supply (included): 120 VAC, 60 Hz, 0.18 amp; 230 VAC, 50/60 Hz, 0.125 amp Allows recharging during ventilator operations/standby Pneumatic Power Internal: 40 to 70 psig (276 to 483 kPa) blended Air/O2; Bleed flow: 6 Lpm (maximum)
Languages:  English, Russian, Japanese
Alarms:  General System Failure; Additional General System Alarms; Low Peak Pressure; Low PEEP; Apnea (optional)
Alarm Controls:  Low Pressure Alarm Delay: 0 to 40 seconds; Proximal Airway Pressure: -20 to 100 cmH20
Non-Adjustable Alarms:  High Pressure: preset at 70 cmH20; Low Pressure: preset at 2.5 cmH20
AC Power and Battery Indicators:  Glows amber with external AC power
Other Specifications:  Other indicators: Battery Low Alarm: flashes red when charging is required; Inspiratory LED: glows amber during mandatory inspiratory phase; Battery Charging: indicates battery charging status

Related Info 

Distributor | Educator | Manufacturer |Rental Agent | Service Center |
Parts Available | Product Markets | Service Centres |

Manufacturer:

Puritan Bennett Corporation
2200 Faraday Avenue
Carlsbad, CA
USA