| Hamilton AMADEUSFT Ventilator | |||
Basic Description | |||
| Manufacturer: | Hamilton Medical AG | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) | ||
| Patient Population Range: | Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Pressure-controlled modes --- SIMV (spontaneous breaths); Flow-controlled modes --- ASSIST/CONTROL; SIMV (mandatory breaths); PCV-CMV; PCV-SIMV Time-cycled modes --- ASSIST/CONTROL; SIMV; PCV-CMV; PCV-SIMV; Apnea backup ventilation | ||
| Waveforms: | Pressure: Rectangular; Flow: Rectangular, Descending Ramp | ||
| Operating Range of Controls: | (S)CMV Frequency: 5 to 120 b/min; SIMV Frequency: 0.5 to 60 b/min; Tidal Volume: 20 to 2000 ml; I:E Ratio: 1:9 to 4:1; Pause: 0 to 8.4 s; Flow: Up to 180 l/min, 2 user-selectable patterns; Pressure Trigger: Off or 1 to 10 cmH2O below PEEP; Flow Trigger: Off or 3 to 15 l/min; Expiratory Trigger Sensitivity: 12.5 to 37.5%; Peak Pressure: 110 cmH2O; PEEP/CPAP: 0 to 50 cmH2O; Pressure Support (Pinsp): 0 to 100 cmH2O; PCV Pressure: 5 to 100 cmH2O; Oxygen: 21 to 100%; Sigh: Every 100 breaths, 50% higher than set tidal volume | ||
| Other Characteristics: | Special functions: Manual breath, apnea backup ventilation, oxygen flush, sigh, hold, pediatric applications; Options: Auxiliary tracheal pressure measurement, medication nebulizer; Monitors:rate, inspiratory peak flow, expired tidal volume, expired minute volume, expired minute volume, maximum pressure, mean pressure (option), autoPEEP, online pressure, static lug-thorax compliance, inspiratory airflow resistance, inspired oxygen concentration, patient inspiratory time (option), I:E ratio (option), leakage volume (option) Special functions: Manual breath, apnea backup ventilation, oxygen flush, sigh, hold, pediatric applications; Options: Auxiliary tracheal pressure measurement, medication nebulizer | ||
Characteristics (Non - Technical) | |||
| Approvals: | CE, CSA, ORKI, COTM, FDA | ||
Specifications | |||
| Alarms: | High rate: 20 to 130 b/min; High pressure: 10 to 110 cmH2O; Low/high minute volume: 0.2 to 50 l/min; Low/high oxygen concentration: 18 to 100%; Apnea time: 20 or 40 s; Disconnection: 2 breaths; Loss of PEEP: 10 s; Flow out of range: > 180 l/min; Gas supply: < 200 kPa (29 psi); Others: User alarms, Flow Sensor alarms, power, inoperative | ||
| Interface Ports: | digital RS 232C, analog, remote nurse call (option), Graphics LEONARDO Advanced Graphics Monitoring (option) | ||
Monitored Variables : | |||
| Pressure | Airway pressure (electronic bar graph): -10 to 110 cmH2O; PEEP/CPAP | ||
| Respiratory Rate | Total rate | ||
| Tidal Volume | Expired for 8 consecutive breaths | ||
| Minute Volume | Expired | ||