| Dräger Evita 1 Ventilator | |||
| (Discontinued) | |||
Basic Description | |||
| Manufacturer: | Dräger Medical, Inc. | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) | ||
| Patient Population Range: | Adult | ||
Characteristics | |||
| Modes--Invasive: | Pressure-controlled modes --- Spontaneous Time-cycled modes --- MMV Combined modes --- CMV: with sensitivity on, breaths are flow-controlled, with sensitivity off, spontaneous breaths are pressure controlled; SIMV: mandatory breaths are flow-controlled, spontaneous breaths are pressure-controlled | ||
| Waveforms: | Pressure: Rectangular, Adjustable Slope; Flow: Rectangular | ||
| Operating Range of Controls: | Sensitivity: 0.5 to 5.0 cmH2O below the baseline pressure (0.7 cmH2O below baseline pressure during SIMV); Ventilator frequency: CMV: 5 to 60 cycles/min; SIMV: 0.5 to 20 cycles/min; Inspiratory pressure: Without pressure support: 0 to 35 cmH2O; With pressure support: pressure limit is 0 to 80 cmH2O above the PEEP/CPAP; Pressure control: 10 to 100 cmH2O; Peak inspiratory flow: 6 to 120 L/min; Tidal volume (if inspiration is flow-controlled): 100 to 2000 mL; I:E ratio: 1:5 to 4:1; Manual breath/inspiratory hold: up to 15 seconds; Intermittent PEEP (2 breaths every 3 minutes): 0 to 35 cmH2O | ||
| Triggering: | Pressure, time, manual | ||
| Graphic Display: | Waveforms for flow and pressure | ||
| Other Characteristics: | If the pressure control setting is below the PIP but above the plateau pressure, the ventilator is designed to maintain the set tidal volume and I:E ratio. Other monitored variables: Inspired gas temperature; Measured oxygen concentration; Compliance; Resistance | ||
Specifications | |||
| Front Panel Controls: | Mode of ventilation; Pressure-triggering threshold; Pressure-limiting threshold; Pressure-cycling threshold; PEEP/CPAP; Tidal Volume; Peak inspiratory flow rate; Pressure support flow; Flow waveform; Ventilatory frequency; I:E ratio | ||
| Power Requirements: | Electrical: 120 VAC, 60 Hz; Pneumatic: 40 to 87 psig | ||
| Alarms: | Power failure; Low gas pressure; Fail to cycle; Mixer inoperative; O2 measurement inoperative; Expiration valve inoperative; Pressure measurement inoperative; Fan defective; Device failure; Apnea; High inspiratory pressure; Low inspiratory pressure; Low oxygen concentration; High oxygen concentration; High inspired gas temperature | ||
| Alarm Controls: | Low/high minute volume: 0 to 41 L/min; High ventilatory frequency: 5 to 80 cycles/min | ||
| Non-Adjustable Alarms: | Power failure; Low gas pressure; Fail to cycle; Mixer inoperative; O2 measurement inoperative; Expiration valve inoperative; Pressure measurement inoperative; Fan defective; Device failure; Apnea; High inspiratory pressure; Low inspiratory pressure; Low oxygen concentration; High oxygen concentration; High inspired gas temperature | ||
| Interface Ports: | RS 232C | ||
Monitored Variables : | |||
| Pressure | Peak, plateau, mean, end expiratory | ||
| Respiratory Rate | Set | ||
| Tidal Volume | Set | ||
| Minute Volume | Spontaneous expired | ||