Dräger Evita 1 Ventilator
(Discontinued)
   

Basic Description

 
 
Manufacturer:  Dräger Medical, Inc.
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Patient Population Range:  Adult
 

Characteristics

 
Modes--Invasive:  Pressure-controlled modes --- Spontaneous
Time-cycled modes --- MMV
Combined modes --- CMV: with sensitivity on, breaths are flow-controlled, with sensitivity off, spontaneous breaths are pressure controlled; SIMV: mandatory breaths are flow-controlled, spontaneous breaths are pressure-controlled
Waveforms:  Pressure: Rectangular, Adjustable Slope; Flow: Rectangular
Operating Range of Controls:  Sensitivity: 0.5 to 5.0 cmH2O below the baseline pressure (0.7 cmH2O below baseline pressure during SIMV); Ventilator frequency: CMV: 5 to 60 cycles/min; SIMV: 0.5 to 20 cycles/min; Inspiratory pressure: Without pressure support: 0 to 35 cmH2O; With pressure support: pressure limit is 0 to 80 cmH2O above the PEEP/CPAP; Pressure control: 10 to 100 cmH2O; Peak inspiratory flow: 6 to 120 L/min; Tidal volume (if inspiration is flow-controlled): 100 to 2000 mL; I:E ratio: 1:5 to 4:1; Manual breath/inspiratory hold: up to 15 seconds; Intermittent PEEP (2 breaths every 3 minutes): 0 to 35 cmH2O
Triggering:  Pressure, time, manual
Graphic Display:  Waveforms for flow and pressure
Other Characteristics:  If the pressure control setting is below the PIP but above the plateau pressure, the ventilator is designed to maintain the set tidal volume and I:E ratio. Other monitored variables: Inspired gas temperature; Measured oxygen concentration; Compliance; Resistance

Specifications

Front Panel Controls:  Mode of ventilation; Pressure-triggering threshold; Pressure-limiting threshold; Pressure-cycling threshold; PEEP/CPAP; Tidal Volume; Peak inspiratory flow rate; Pressure support flow; Flow waveform; Ventilatory frequency; I:E ratio
Power Requirements:  Electrical: 120 VAC, 60 Hz; Pneumatic: 40 to 87 psig
Alarms:  Power failure; Low gas pressure; Fail to cycle; Mixer inoperative; O2 measurement inoperative; Expiration valve inoperative; Pressure measurement inoperative; Fan defective; Device failure; Apnea; High inspiratory pressure; Low inspiratory pressure; Low oxygen concentration; High oxygen concentration; High inspired gas temperature
Alarm Controls:  Low/high minute volume: 0 to 41 L/min; High ventilatory frequency: 5 to 80 cycles/min
Non-Adjustable Alarms:  Power failure; Low gas pressure; Fail to cycle; Mixer inoperative; O2 measurement inoperative; Expiration valve inoperative; Pressure measurement inoperative; Fan defective; Device failure; Apnea; High inspiratory pressure; Low inspiratory pressure; Low oxygen concentration; High oxygen concentration; High inspired gas temperature
Interface Ports:  RS 232C

Monitored Variables :

Pressure Peak, plateau, mean, end expiratory
Respiratory Rate Set
Tidal Volume Set
Minute Volume Spontaneous expired

Related Info 