| Bear 1000 Ventilator: Intermediate Model | |||
Basic Description | |||
| Manufacturer: | Bear Medical Systems Incorporated | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) CPAP | ||
| Patient Population Range: | Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- ASSIST/CMV; SIMV/CPAP (PSV) Pressure-controlled modes --- PRESSURE CONTROL Combined modes --- PRESSURE AUGMENT (pressure/volume control) | ||
| Waveforms: | Square, Descending, Sine | ||
| Operating Range of Controls: | Tidal Volume: 0.10 to 2.00 L, 0.03 to 2.00 L; Rate: 0, 0.5, 1 to 120 BPM; Peak Flow: 10 to 150 LPM, 5 to 150 LPM; %O2: 21 to 100%; Pressure Support: 0 to 80 cmH2O; Assist Sensitivity: 0.2 to 5 cmH2O; PEEP/CPAP: 0 to 50 cmH2O; Inspiratory Pause: 0 to 2 seconds; MMV Level: 0 to 50 LPM; Compliance Comp.: 0 to 7.5 ml/cmH2O; Inspiratory Time: 0.1 to 5 seconds; Pressure Slope: -9 to +9 and P-9 to P+9; Waveforms: Square, Decelerating, Sine; Manual Breath: x1; Manual Inspiratory Pause: 2 Seconds Maximum; Expiratory Hold: 9 Seconds Maximum; 100% O2: On/Off-3 Minutes Auto Off; Sighs:On/Off (Every 100 Breaths at 150% of set Tidal Volume); Pressure Augmentation: On/Off; Nebulizer: On/Off-30 Minutes Auto Off; Panel Lock: On/Off; Set Knob: No End Stop; Inspiratory Pressure: 0 to 80 cmH2O | ||
| Graphic Display: | Waves Page Flow, Volume, and Pressure (AW) waveforms with user adjustable waveforms scales and sweep speed Loops Page Pressure/Volume AW (airway pressure) Flow Volume Monitoring Data VTI, VTE, PIFR, PIP, PEEP Mechanics Data CRS: Static Compliance 0 to 300 ml/cmH2O RRS: Resistance of the Respiratory System 0 to 100 cmH2O/L/sec | ||
| Size: | 23"W x 13.5" D x 15"H | ||
| Weight: | 54 lbs | ||
Characteristics (Non - Technical) | |||
| Approvals: | IEC UL | ||
Specifications | |||
| Power Requirements: | 95 - 135V, 60 Hz (U.S./ International); 80 - 110V, 50/60 Hz (International); 176 - 242V, 50/60 Hz (International); 192 -264V, 50/60 Hz (International) | ||
| Alarm Controls: | Total Minute Volume: High 0 to 80 LPM, Low 0 to 50 LPM; Total Breath Rate: High 0 to 155 BPM, Low 1 to 99 BPM; Peak Inspiratory Pressure: High 0 to 120 cmH20, Low 3 to 99 cmH2O; Baseline Pressure: High 0 to 55 cmH2O, Low 0 to 50 cm H20 | ||
| Non-Adjustable Alarms: | Time/I:E Limit; Run Diagnostics; Gas Supply; Failed to Cycle; I:E Override | ||
Monitored Variables : | |||
| Pressure | 0 to 140 cmH20 (Peak, Mean, and Plateau) | ||
| Respiratory Rate | 0 to 155 BPM | ||
| Tidal Volume | 0.00 to 9.99 L | ||
| Minute Volume | 0.0 to 99.9 LPM | ||
Related Info | |||