| Bird V.I.P. Bird Infant & Pediatric Ventilator | |||
Basic Description | |||
| Manufacturer: | Bird Products Corporation | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) | ||
| Patient Population Range: | Neonatal Pediatric | ||
Characteristics | |||
| Modes--Invasive: | Pressure-controlled modes --- SIMV (spontaneous breaths); ASSIST/CONTROL (inspiration) (optional)) Flow-controlled modes --- ASSIST/CONTROL; SIMV (mandatory breaths) Time-cycled modes --- (S)IMV/CPAP; ASSIST/CONTROL (optional) | ||
| Waveforms: | Pressure: Rectangular; Flow: Rectangular | ||
| Operating Range of Controls: | Tidal Volume: 20 to 995 ml; Inspiratory Time: 0.10 to 3.00 sec.; Breath Rate: 0 to 150 bpm; Flow -- Time Cycled Flow: 3 to 40 lpm; Flow -- Volume Cycled: 3 to 100 lpm; High Pressure Limit -- Time Cycled: 3 to 80 cmH2O; High Pressure Limit -- Volume Cycled: 3 to 120 cmH2O; PEEP/CPAP: 0 to 24 cmH2O, Off; Assist Sensitivity:iFlow Triggered 0.2 to 5.0 lpm, Off; Pressure Triggered: 1 to 20 cmH2O, Off; Pressure Support: 1 to 50 cmH2O, Off; Termination Sensitivity: Time Cycled, Flow Synchronized 5% to 25% of Peak Flow, Off; Manual Breath: x 1; Oxygen Percent: 21% to 100%; Over-Pressure Relief: 0 to 130 ± 10 cmH2O | ||
| Triggering: | Pressure, Time, Manual | ||
| Power: | AC 110V/60Hz | ||
| Other Characteristics: | Also displays inspiratory time, I:E ratio, patient effort, demand | ||
Specifications | |||
| Front Panel Controls: | Power; AC/DC; Modes: Volume Cycled: SIMV/CPAP; ASSIST/CONTROL; Time Cycled: (S)IMV/CPAP; ASSIST/CONTROL; Tidal Volume; Inspiratory Time; Breath Rate; Flow; High Pressure Limit; PEEP/CPAP; Assist Sensitivity; Pressure Support; Termination Sensitivity; Oxygen Percent; Over-Pressure Relief; High Pressure Limit; Low Peak Pressure; Low PEEP/CPAP Pressure; High/Prolonged Pressure; Apnea Interval; Low Inlet Gas Pressure | ||
| Power Requirements: | Electrical: 115 VAC, 60 Hz or optional European voltages; Pneumatic: External compressed gas at 40 to 75 psig | ||
| Alarm Controls: | High Pressure Limit: 3 to 120 cmH2O; Low Peak Pressure: 3 to 120 cmH2O, Off; Low PEEP/CPAP Pressure: -9 to + 24 cmH2O; High/Prolonged Pressure: a) Time Cycled: High Pressure Limit + 10 cmH2O, Auto set; b) Volume Cycled: 3 to 120 cmH2O; c) All Modes (Prolonged Pressure): PEEP/CPAP + 6 cmH2O; Apnea Interval 20, 40 and 60 sec. | ||
| Non-Adjustable Alarms: | Low Inlet Gas Pressure: System Pressure <22.5 or 27.5< PSIG Red Light Indicator Lamp; Circuit Fault: Red Light Indicator Lamp | ||
| Mounting Options: | Wall mount, stand | ||
Monitored Variables : | |||
| Pressure | PIP, mean airway pressure, airway pressure (aneroid gauge) | ||
| Respiratory Rate | |||
Related Info | |||