Hamilton ARABELLA Active Nasal CPAP System
   

Basic Description

 
 
Manufacturer:  Hamilton Medical AG
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Subacute
CPAP
Patient Population Range:  Neonatal
 

Characteristics

Operating Range of Controls:  Flow 0 to 10 l/min; Oxygen 21 to 100 %; Monitoring; NCPAP 0 to 12 cmH2O; Oxygen 21 to 100%
Size:  21.4 x 17.5 x 14.5 cm (8.4 x 6.9 x 5.7 in.)
Weight:  7.1kg (15.6 lbs)

Characteristics (Non - Technical)

Approvals:  IEC 60601-1/ EN60601-1 IEC 60601-1-2/ EN60601-1-2, CE
Other Comments:  Formerly known as the ALADDIN II

Specifications

Power Requirements:  Electrical power: 110 to 240 VAC, 50/60 Hz, 22VA; Air/oxygen supplies: 207 to 510 kP a (2 to 5 bar, 30 to 75 psi)
Alarms:  High pressure: 3 cmH2O above set NCPAP level; Low pressure: 2 cmH2O below set NCPAP level; High and low Oxygen: +/-5% of set Oxygen level; Alarm-silence button: 3min alarm silence; Internal pop-off valve: 150 cm H2O; Safety limit valve and alarm (reinitiates gas flow after 3s): terminates flow if pressure exceeds 11mbar; Gas supply: Pressure difference between air and oxygen is > 200kPa (2 bar, 29 psi)
Other Specifications:  Operating Temperature: 10 to 40 C at 0 to 75 reIH, noncondensing; Storage Temperature: -10 to 60 C at 0 to 75 reIH, noncondensing; Atmospheric Pressure: 700 to 1060 hP a (10 to 15 psi)