Bear Cub (BP 2001) Ventilator
(Discontinued)
   

Basic Description

 
 
Manufacturer:  Bear Medical Systems Incorporated
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Transport
CPAP
Patient Population Range:  Neonatal
 

Description

 
Features:  Adjustable Pressure Limit to prevent high peak airway pressures. Volume Limit ™ to prevent high tidal volume delivery--limits inspiratory tidal volume and has clinician-controlled audible limit alarms. Flow Cycling for improved synchrony--delivers the appropriate inspriatory time for the infant's respiratory dynamics.
 

Characteristics

 
Modes--Invasive:  Pressure-controlled modes --- CMV/IMV
Supplemental modes --- CPAP (not controlled)
Waveforms:  Pressure: Exponential; Flow: Rectangular
Operating Range of Controls:  Oxygen%; 21% to 100%; PEEP/CPAP: 0 to 20 cmH2O (uncalibrated knob); Pressure Limit: 0 to 72 cmH2O; Flow: 3 to 30 L/min; Inspiratory Time: 0.1 to 3.0 seconds; Ventilator Rate: 1 to 150 breaths/minute; Lamp Battery Test, CPAP, CMV/IMV; Ventilator Rate: 1 to 150 cycles/minute
Power: 
Other Characteristics:  The Bear Cub also monitors inspiratory time, exhalation time, and I:E ratio.

Specifications

Front Panel Controls:  Loss of PEEP/CPAP; Low inspiratory pressure alarm; Oxygen%; PEEP/CPAP; Pressure Limit; Flow; Visual Reset; Alarm Silence; Inspiratory Time; Ventilator Rate; Lamp Battery Test, CPAP, CMV/IMV; Ventilator Rate; Toggle for ventilator rate range; Manual Breath
Power Requirements:  Electrical power: 117 VAC, 60 Hz; Pneunatic power: Air at 15 to 75 psig; Oxygen at 30 to 75 psig
Alarm Controls:  Low inspiratory pressure: off to 50 cmH2O; Loss of PEEP/CPAP: off to 20 cmH2O
Non-Adjustable Alarms:  Source gases below 22.5 +/- 2.5 psig; Prolonged inspiratory pressure: proximal airway pressure > 10 cmH2O above the loss of PEEP/CPAP alarm for > 3.5 seconds; I:E ratio > 3:1; Rate/time incompatibility; Ventilator inoperative: Electrical power interruption; Failure to cycle; Control panel malfunction; Prolonged solenoid "on" time; Inspiratory time variation > -/24% of setting; Timing circuit failure

Monitored Variables :

Pressure
Respiratory Rate

Related Info 

Distributor |Educator | Manufacturer |Rental Agent |Service Center |
Parts Available | Product Markets | Service Centres |

Distributor:

Freedom Medical
208 Carter Drive
West Chester, PA



Manufacturer:

Bear Medical Systems Incorporated
1100 Bird Center Drive
Palm Springs, CA
USA



Rental Agent:

Freedom Medical
208 Carter Drive
West Chester, PA