Bear 5 Ventilator
(Discontinued)
   

Basic Description

 
 
Manufacturer:  Bear Medical Systems Incorporated
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Patient Population Range:  Adult
 

Characteristics

 
Modes--Invasive:  Pressure-controlled modes --- SIMV/IMV & AMV (spontaneous breaths); CPAP
Flow-controlled modes --- CMV; ASSIST CMV; SIMV/IVM & AMV (mandatory breaths)
Time-cycled modes --- TIME CYCLE
Waveforms:  Pressure: Rectangular; Flow: Rectangular, Ascending Ramp, Descending Ramp, Sinusoidal
Operating Range of Controls:  Tidal Volume: 50 to 2000 mL; Normal Rate: 0 to 150 cylcles/min; Peak Flow: 5 to 150 cmH2O L/min; %O2: 21% to 100%; PEEP CPAP (baseline pressure): 0 to 50 cmH2O with pressure support; 0 to 72 cmH2O above PEEP/CPAP without pressure support; Inspiratory Pause: 0 to 2.0 seconds; Assist Sensitivity: 0.5 to 5.0 cmH2O below the baseline pressure; Inspiratory Time: 0.1 to 3.0 seconds; Continuous Flow: 5 to 40 L/min; Sigh Volume: 65 to 3000 mL; Sigh Rate: 2 to 60 sighs/minute; Multi Sigh: sighs in groups of 2 or 3
Power:  AC 110V/60Hz

Specifications

Front Panel Controls:  Alarm Silence; Alarm Reset; Test; Power; Modes: CMV; Assist CMV; SIMV-IMV; PEEP; AMV; Time Cycle; Panel Lock; Tidal Volume; C Comp; Normal Rate; Peak Flow; %O2; PEEP CPAP; Inspiratory Pause; Assist Sensitivity; Pressure Support; Minute MV; Inspiratory Time; Cont Flow; Pressure Relief; Sigh Volume; Sigh Rate; Multi Sigh; Sigh Off; Sigh On; Manual Sigh; Manual Breath; 100% O2; Rectangular Waveform; Ascendending Ramp Waveform; Descending Ramp Waveform; Sinusoidal Waveform; Keypad
Power Requirements:  Electrical: 115 VAC, 60 Hz; Pneumatic: External: 30 to 100 psig; Internal Compressor: 9.5 psig
Alarms:  Alarm thresholds may be set manually or by allowing the ventilator to provide automatic ranges for specific modes of ventilation.
Alarm Controls:  Low PEEP/CPAP: 0 to 50 cmH2O; High PEEP/CPAP: 0 to 55 cmH2O; Low mean airway pressure: 0 to 75 cmH2O; High mean airway pressure: 0 to 75 cmH2O; Low inspiratory pressure: 0 to 140 cmH2O; High peak normal pressure: 0 to 140 cmH2O; High peak sigh pressure: 0 to 140 cmH2O; Low mandatory volume: 0 to 3000 mL; Low spontaneous volue: 0 to 3000 mL; Low exhaled minute volume: .3 to 40 L/min; High exhaled minute volume: 1.0 to 80 L/min; Low breath rate: 3 to 155 cycles/min; High breath rate: 3 to 155 cycles/min; Low inspiratory time: 0.05 to 3.00 seconds; High inspiratory time: 0.10 to 3.20 seconds
Non-Adjustable Alarms:  Electrical power < 102 volts; Pneumatic power < 27 psig; Flow sensor malfunction; Prolonged high airway pressure; Airway pressure at the maximum working pressure; Open temperature control chamber door; I:E ratio

Monitored Variables :

Pressure -10 to 150 cmH2O (gauge)

Related Info 

Distributor | Educator | Manufacturer |Rental Agent | Service Center |
Parts Available | Product Markets | Service Centres |

Manufacturer:

Bear Medical Systems Incorporated
1100 Bird Center Drive
Palm Springs, CA
USA