Bear 1 Ventilator
(Discontinued)
   

Basic Description

 
 
Manufacturer:  Bear Medical Systems Incorporated
Ventilator Type:  Positive Pressure
Invasive
Conventional
Ventilator Class:  Intensive Care Unit (ICU)
Transport
CPAP
Patient Population Range:  Pediatric
Adult
 

Characteristics

 
Modes--Invasive:  Volume-controlled modes --- CONTROL; ASSIST/CONTROL; SIMV (mandatory breaths)
Pressure-controlled modes --- SIMV (spontaneous breaths); CPAP
Waveforms:  Continuously adjustable from rectangular to descending ramp (truncated exponential decay)
Operating Range of Controls:  Tidal Volume: 100 to 2000 mL; Normal Rate: 0.5 to 6 bpm or 5 to 60 bph; Normal Pressure Limit: 0 to 100 cmH2O; Sigh Volume 150 to 3000 mL; Sigh Rate: 2 to 60 sighs per hour; Sigh Pressure Limit: 0 to 100 cmH2O; Assist: (less/more sensitivity); Oxygen: 21% to 100%; Peak Flow: 20 to 120 L/min; Inspiratory Pause 0 to 2 seconds; PEEP: 0 to 30 cmH2O
Triggering:  Pressure; Volume Flow; Time; Manual

Specifications

Front Panel Controls:  Power Loss Alarm Test; Mode; Single Breath; Tidal Volume; Normal Rate; Toggle for Range of Normal Rate; Normal Pressure Limit; Standby; Pressure (toggles machine/proximal); Minute Volume Accumulate; Sigh (off/1/2/3); Single Sigh; Sigh Volume; Sigh Rate; Sigh Pressure Limit; Wave Form; Assist (less/more sensitivity); 1:1 Ratio Limit (off/on); Oxygen; Peak Flow; Inspiratory Pause; Nebulizer (off/on); PEEP
Power Requirements:  Electrical: 115 VAC, 60 Hz; Pneumatic: External: 25 to 100 psig; Internal Compressor: 9.5 psig
Alarms:  General System Failure; Inverse I:E Ratio; High Peak Pressure; Low Peak Pressure; Low PEEP; Low Tidal Volume; Apnea
Alarm Controls:  Peak Inspiratory Pressure (PIP): 0 to 100 cmH2O (same as Normal Pressure Limit); Low Inspiratory Pressure: 0 to 50 cmH2O; Low PEEP/CPAP: 0 to 30 cmH20; Minimum Exhaled Volume: 0 to 2 L/min (three consecutive breaths)
Non-Adjustable Alarms:  Visible and audible alarm: Interrupted electrical power; Interrupted air supply; Interrupted oxygen supply; Apnea; Flow sensor malfunction; I:E ratio > 1:1; Inspiratory flow > 20 L/min for more than 9 to 12 seconds; Flashing LED: I:E ratio < 1:10

Monitored Variables :

Pressure Proximal Airway Pressure: -10 to 100 (gauge)
Respiratory Rate
Tidal Volume
Minute Volume