| Newport E100M Ventilator | |||
Basic Description | |||
| Manufacturer: | Newport Medical Instruments | ||
| Ventilator Type: | Positive Pressure Invasive Conventional | ||
| Ventilator Class: | Intensive Care Unit (ICU) Transport Subacute CPAP | ||
| Patient Population Range: | Neonatal Pediatric Adult | ||
Characteristics | |||
| Modes--Invasive: | Volume-controlled modes --- ASSIST/CONTROL; SIMV Pressure-controlled modes --- ASSIST/CONTROL; SIMV; CPAP | ||
| Operating Range of Controls: | %O2: 21 to 100%; V(Mandatory Flow): 1 to 100 L/min; Inpiratory Time: 0.1 to 3.0 sec; Freq/Rate: 1 to 120 b/min; PEEP/CPAP: 0 to +25 cmH2O; Pressure Limit: 0 to 120 cmH2O; Sensitivity: -10 to +25 cmH2O; Time-Limited Demand Flow: 0 to 1.0 sec; | ||
| Size: | 24.1 x 16.5 x 26.7 (cm) | ||
| Weight: | 8.2 (kg) | ||
| Other Characteristics: | Battery 6-8 hours typical, 4 hours @ max load. | ||
Characteristics (Non - Technical) | |||
| Approvals: | CE,NSAI,ISO | ||
| Other Comments: | Preventative maintenance every 3,000 hours or once/year, overhaul every 15,000 hours or every 5 years. Contact Newport for distributors. | ||
Specifications | |||
| Alarms: | High Pressure: Audible and Visual; Low Pressure: Audible and Visual; Low Battery: Audible and Visual; APNEA: Audible and Visual; Low Baseline (P min): Audible and Visual; High Baseline (P max/P min): Audible and Visual; Fault: Audible; Gas Supply Source Failure: Audible; Power Failure: Audible | ||
| Other Specifications: | Battery duration 6 to 8 hours | ||
Monitored Variables : | |||
| Pressure | |||
| Tidal Volume | |||